U. 3. WAS CEPT, TECHNICAL MANUAL 8-627 | DIATHERMY APPARATUS SHORT WAVE, COMPLETE, 110 VOLT. 60 CYCLE, ITEM 7105505 mmi WAR DEPARTMENT TEINNICAl MANUAL DIATHERMY APPARATUS, SHORT WAVE, COMPLETE, 110 VOLT, 60 CYCLE, ITEM 7105505 WAR DEPARTMENT NOVEMBER 1944 WAR DEPARTMENT TECHNICAL MANUAL TM 8-627 DIATHERMY APPARATUS, SHORT WAVE, COMPLETE, 110 VOLT, 60 CYCLE, ITEM 7105505 WAR DEPARTMENT N 0 EE M B E R 1944 United States Government Printing Office Washington ■ 1944 WAR DEPARTMENT, Washington 25, D. C., 3 November, 1944. TM 8-627, Diathermy Apparatus, Short Wave, Complete, 110 Volt 60 Cycle, Item 7105505, is published for the information and guidance of all concerned. [A. G. 300.7 (4 Oct 44).] By order of the Secretary of War: G. C. MARSHALL, Chief of Staff. Official : J. A. ULIO, Major General, The Adjutant General. Distribution : Armies (2) ; Corps (2) ; Dcpts (10) ; Sv C (5) ; ID 8 (4) ; IB 8 (1); IR 8 (3) (2) (1) ; 1C 8 (2) ; Base Comds (5); Numbered Air Forces (5); Med Deps, Continental (25) except St. Louis Med Dcp (50), Med Sec ASF Dep (10), T of Opn (10); PE (2). ID 8: T/O & E 8-550. IB 8: T/O & E 8-611; 8-610. IR 8 (3) : T/O & E 8-581; 8-580, Evac Hosps. IR 8 (2) : T/O & E 8-560, 8-510; 8-590; 8-572S. IR 8 (1) : T/O & E 8-580, Gonval Hosps. IC 8: T/O & E 8-187, 8-667. For explanation of symbols, see FM 21-6. CONTENTS CHAPTER 1. INTRODUCTION. Section I. General. Paragraph Page Scope 1 1 II. Description and data. Description 2 1 Data 3 1 CHAPTER 2. OPERATING INSTRUCTIONS. Section I. General. Scope 4 3 II. Service upon receipt of equipment. Unpacking 5 3 Assembling 6 3 Conditioning rectifier tubes 7 7 III. Controls and instruments. Controls 8 7 Instruments 9 8 IV. Operation. Procedure for operation 10 10 CHAPTER 3. MAINTENANCE INSTRUCTIONS. Section I. General. . Scope . 11 11 II. Lubrication. Blower motor 12 11 III. Preventive maintenance services. Daily 13 11 Three months 14 11 IV. Trouble shooting. White pilot light not lit when main switch is turned on. . 15 11 Plate current too high 16 11 Heat not reaching patient but voltmeter registering. . . 17 12 V. Maintenance operations. Dusting. . 18 12 “Trimmer” adjustment 19 12 Appendix. List of all service parts 15 CHAPTER 1 INTRODUCTION Section I. GENERAL 1. SCOPE, a. These instructions are published for the information and guidance of all personnel charged with the operation and maintenance of Diathermy Apparatus, Medical Department item No. 7105505 (fig. 1) in the field. They contain information on the operation and maintenance of the equipment with illustrations showing the various parts referred to in the instructions. They are arranged in three parts: Chapter 1—Intro- duction; Chapter 2—Operating Instructions; Chapter 3—Maintenance Instructions. b. A list of all service parts is included in the appendix. Those parts keyed with an asterisk are spare parts regularly stocked for issue. Those parts not keyed with an asterisk are available through special purchase only. Section II. DESCRIPTION AND DATA 2. DESCRIPTION. The Burdick Model SWDX-80 Crystal Controlled Short Wave Diathermy supplies high frequency energy to either an in- ductance cable applicator, or a drum applicator, at a frequency of 13.660 megacycles (approximately 22 meters). Frequency control devices within the unit actuated by a piezoelectric quartz crystal maintain the frequency within a band width of plus or minus 0.05 percent of 13.660 me, regard- less of how the controls are operated or the applications to patient are made. 3. DATA. a. Performance. This machine is designed to operate on 100-130 volt, 50-60 cycles, alternating current. b. Manufacturer. This machine is manufactured by The Burdick Corporation, Milton, Wisconsin. Med. Dept. No. Nomenclature I 7R08170 Screw, 1/4-20X3/4 Inch, Thumb, Brass. Figure 1. Item No. 7105505 Diathermy apparatus. CHAPTER 2 OPERATING INSTRUCTIONS Section I. GENERAL 4. SCOPE. The professional use of the Diathermy Apparatus is the re- sponsibility of the Medical Officer. However, chapter 2 contains in- formation on assembling and the mechanics of operating the equipment. Section II. SERVICE UPON RECEIPT OF EQUIPMENT 5. UNPACKING. This unit is shipped in two cases, one of which contains the cabinet assembly and the other contains the arm, arm holder, treatment drum, cables and tubes. Open the case containing the cabinet and remove all supporting braces, screws and bolts. Remove packing materials and lift apparatus from the case. Open the other case to provide access to the contents. 6. ASSEMBLING, a. Remove both front and rear doors of cabinet. Front door is removed by taking out the two knurled head screws, 7R08170 (fig. 1), at bottom edge of door and pulling bottom of door out and down. Rear door is removed by taking out the 10 screws, SR00575 (fig. 2), which hold it on. b. Carefully inspect the interior for parts loosened or broken in ship- ment. Remove all strings used to tie down tube connectors. c. Attach the arm holder, 7R08252 (figs. 2 and 3), to cabinet in the position shown in figure 3. Bfefore tightening the clamp, check the vertical position to be sure it registers with the notch in rear door. d. Install the following tubes; Two FP-265 final amplifiers, 7R08104 (see fig. 3). One GL-814 exciter amplifier, SR00061 (see fig. 4). One RCA-6-L6 Metal, oscillator, SR00062 (see fig. 4). One RCA-5Y3GT exciter power supply rectifier, SR00064 (fig. 4). Two RCA-866A high voltage supply rectifier, SR00063 (fig. 4). (1) Insert the FP-265 tubes, 7R08104 (fig. 3), in the sockets at either side of final amplifier tank. These tubes fit in the sockets located directly behind the shield plates, 7R08280 (see fig. 3). Carefully rotate the tubes so the tube base pins will register with the socket holes, then push tube down as far as it will go. Attach the plate clips, 7R08188 (fig. 3), on top caps of tubes. The clips should not be snapped into position from the side, but forced gently over the ends of top caps on tubes. These leads are very short and care must be exercised not to damage them or put a strain on the tank coil when placing the clips on the tube caps. Med. Dept. No. 1. SR00426 2. 7R08250 3. SR00575 Nomenclature Receptacle, Male, Twist Lock, Three Wire, Hubbell No. 7556. Plate, Cover, ‘‘Trimmer’’ Condenser. Screw, 6 32 XI/2-Inch, FL H. M. Med. Dept. No. 4. 7R08148 5. 7R08278 6. 7R08252 Nomenclature Drum, Treatment, Complete. Arm, Treatment Drum, Adjustable. Holder, Treatment Arm, Complete. 4 Figure 2. Rear View, Diathermy apparatus Med. Dept. No. 1.7R08280 Plate, Shield. 2.7R08104 Tube, Amplifier, High Mu Triode. 3.7R08188 Clip, Plate, Amplifier Tube, Complete. 4.7R08252 Holder, Treatment Arm, Complete. 5. 7R08288 Cover, Relay. Nomenclature Figure 3. Interior view of rear section of Diathermy apparatus. 614880—44—2 Med. Dept. No. 1. SR00061 Tube, Amplifier, Beam Tetrode. 2. SR00062 Tube, Oscillator, Beam Tetrode, Metal. 3. 7R08102 Crystal, Quartz. 4. SR00064 Tube, Rectifier. 5. SR00063 Tube, Rectifier, Mercury. Nomenclature Figure 4. Interior view of front section of Diathermy apparatus, (2) The GL-814 tube is mounted horizontally in a socket on the shield wall of the exciter unit as shown in figure 4. Rotate the tube carefully to locate base pins before pushing tube into socket. Be sure to push the tube into socket as far as possible. Attach plate clip to the top cap on the tube, (3) Insert the RGA-6L6 tube in the socket at the extreme right rear of exciter unit as shown in figure 4. Rotate the tube carefully to locate base pins before pushing tube into socket. (4) Insert the RCA-5Y3GT tube in the socket on the exciter power supply unit, as shown in figure 4. Rotate the tube carefully to locate base pins before pushing tube into socket. (5) Insert the RCA-866A tubes in the two sockets between the exciter power supply unit and the plate transformer, as shown in figure 4. Rotate the tube carefully to locate base pins before pushing tube into socket. Attach the plate clips to the top caps of tubes. (6) Insert the MC-5 Bliley 6830 KG piezoelectric quartz crystal in the socket at the right front to exciter unit, as shown in figure 4. e. Replace both doors. f Attach the treatment drum, 7R08148 (fig. 2), and adjustable arm, 7R08278 (fig. 2), and mount in the arm holder, 7R08252 (see figs. 2 and 3). 7. CONDITIONING RECTIFIER TUBES. When new 866-A mercury recti- fier tubes are first placed in operation, or tubes have been out of their sockets so the mercury has come in contact with either the filament or anode, the filaments should be operated at least 5 minutes with the plate voltage off. This is necessary in order to evaporate the mercury collected on the filament and anode, as otherwise the rectifier tubes will be damaged. To condition the rectifier tubes, start with oscillator switch in OFF posi- tion, turn on main switch, and turn voltage adjuster until pointer of voltmeter is in black area of scale. Allow filaments to run for at least 5 minutes. After at least 5 minutes have elapsed, with power controls and resonance control on ZERO, turn oscillator switch on. The rectifier tubes should have a steady blue color. If there is sparking or flashing in the tubes, the preheating must continue until flashing stops. This can be observed through the vents in front door. After this conditioning treat- ment, the preheating time is automatically maintained at 15 seconds by the time delay relay. This delay is necessary to protect the 866-A mer- cury rectifier tubes from receiving plate voltage before the filaments have reached normal operating temperature. Section III. CONTROLS AND INSTRUMENTS 8. CONTROLS, a. Main switch. The main switch, 7R08204 (fig. 5), turns on current to all tube filaments and exhaust blower. When the switch is in the ON position, the voltmeter is registering and the white pilot light, SR00071 (fig. 5), is lighted. The main switch is the circuit breaker type and takes the place of line fuses in addition to acting as the on-and- off switch. b. Oscillator switch. This switch, 7R08166 (fig. 5), turns on current to the oscillator and amplifier circuits and must be in the ON position before any energy can be produced by the machine. The RED pilot light should light up and the plate current milliammeter indicate when oscillator switch is ON. However, due to the time delay relay, when oscillator switch is turned ON, the RED pilot light will not light and the plate current milliammeter will not indicate until approximately 15 seconds after the main switch is turned ON. Whenever the RED pilot light is out when oscillator switch is ON, it is because the overload relay has tripped. This is reset and power restored by turning power control below ZERO to position marked RESET. c. Voltage adjuster. The voltage adjuster, 7R08282 (fig. 5), is a line voltage compensator by which variations in line voltage are adjusted to provide proper filament and plate voltage. This control and the volt- meter work in conjunction. Should the needle of the voltmeter fall below the black area, turn the voltage adjuster to a higher point. Should the reading be above the black area, turn the voltage adjuster to a lower point. Be sure to maintain the position of the needle in the black area as near the center arrow as possible, under all applications of the unit. Due to line voltage drop, it is usually necessary to turn up the voltage adjuster when treatment is started. d. Resonance control. This control, 7R08284 (fig. 5), tunes the out- put or patient circuit so that it resonates with the final amplifier, an oper- ation necessary to transfer energy to the patient. This control must always be adjusted to give maximum plate current, which is the indication of resonance. The position on the resonance control dial at which resonance occurs depends on the particular application given patient. This control does not act the same as the power control, because after resonance has been reached (maximum plate current) further increase of the resonance control causes a reduction in plate current. It is necessary to adjust resonance control for maximum plate current (resonance) in order for the plate current to be an indication of power output. e. Power control. This control, 7R08286 (fig. 5), is a variable mag- netic coupling of the output (patient) circuit to the final power amplifier, and provides a continuously variable control of the energy reaching the patient, indicated by the plate current millammeter. If the control is turned too high so the plate current exceeds 350 ma, an overload relay trips and shuts off power. Power is restored by turning power control down below ZERO to a position marked RESET. f. Patient’s emergency cut-off. This cord, 7R08176 (fig. 5), to be held by the patient, is attached to the oscillator switch and when pulled from any position it will open the switch. 9. INSTRUMENTS, a. Voltmeter. The purpose of the voltmeter, 7R08130 (fig. 5), is to indicate the voltage which is being supplied to the tube filaments and plates. Proper voltage is being supplied when the meter registers in the black area as near as possible to arrow in the center. This is necessary to the proper operation of the unit and to maintain normal tube life. Control is accomplished by use of the voltage adjuster explained in paragraph 8c. b. Plate current d-c milliammeter. This milliammeter, 7R08I32 (fig. 5), indicates the d-c plate current and is directly proportionate to the radio frequency power which is being drawn from the oscillator by the patient circuit. In other words, this is an arbitrary guide to the amount of energy being generated. In order for the plate current to be a true indication of power output it is necessary that (1) the unit should operate long enough with power control and resonance control on ZERO to allow 8 Figure 5. The control panel. Med. Dept. No. Nomenclature 1 7R08282 Adjuster, Voltage. 2 7R08204 Switch, Voltage Control, Complete. 3 7R08130 Voltmeter, 3 1/2-Inch, Triplett. 4 SR00071 Bulb, Pilot, 6-8 V, Miniature Screw Base. 5 7R08286 Control, Power. 6 SR00071 Bulb, Pilot, 6-8 V, Miniature Screw BBSS 7 7R08132 Milliammeter, 3 1/2- Inch, Triplett. 8 7R08176 Lever, Patient, Pull- Off, Complete. 9 7R08166 Switch, Toggle, Oscillator. 10 7R08284 Control, Resonance. 9 plate current to reach a minimum value before starting treatment (2) resonance control must be tuned for maximum plate current during treatment. c. White pilot light. This light, SR00071 (fig. 5), indicates when main switch is ON and tube filaments and blower are operating. d. Red pilot light. This light, SR00071 (fig. 5), indicates when oscilla- tor switch is ON and unit is generating high frequency energy. If the red pilot is out when oscillator switch is ON, the indication is that the overload relay has tripped. Section IV. OPERATION 10. PROCEDURE FOR OPERATION, a. See that the current supply is 60-cycle (or 50-cycle) a-c of a voltage between 100 and 130 volts. b. The short wire extending from the wall outlet plug on line cable is to be connected to a good ground. If the building wiring has been in- stalled in properly grounded conduit and iron outlet boxes, it is sufficient to attach this ground lead to the outlet box by means of one of the screw plugs supplied with each unit. If the wiring is not in grounded conduit, it will be necessary to extend this ground lead to a water pipe, radiator or some other approved ground connection. This ground wire in the line cord connects to the steel cabinet when the line cable is plugged into cabi- net receptable. Grounding the cabinet is not essential to the proper func- tioning of the unit, but is a safety measure. It prevents any possible danger to patient and operator. c. See that both switches, “Main” and “Oscillator,” are in the OFF position. Turn Power Control and Resonance Control to ZERO. d. Plug line cable into the three-wire receptacle, SR00426 (fig. 2), on the right rear of cabinet near base. e. Turn on main switch and adjust voltage to upper range of black area on voltmeter scale. After about 15 seconds the time delay relay will click and oscillator switch can be turned on. f. Wait several minutes for the plate current to come down to its minimum value, (90 ma or less). This preheating period is necessary to allow the final amplifier to reach operating temperature. g. With the cable or drum applied to patient, and cable jacks plugged into outlets at rear of cabinet marked “Cable” (fig. 2) and after enough time has elapsed for plate current to reach a minimum value, the treat- ment can be started. h. Advance power control to about 1 or 2 and tune to resonance with resonance control. The exact resonance peak is indicated by maximum reading on the plate current milliammeter. Slowly adjust the resonance control to obtain maximum plate current. Rock the control back and forth to find the exact peak. If the resonance control is turned too fast, the resonance peak may be passed over unnoticed. i. Now with the circuit tuned to resonance, advance the power control to give the desired amount of energy to the patient, indicated (relatively) by the plate current. j. When the plate current exceeds 350 ma an overload relay trips and shuts off the high frequency power, indicated by the RED pilot light going out. The relay is reset and power restored by turning power control below ZERO to the position marked RESET. 10 CHAPTER 3 MAINTENANCE INSTRUCTIONS Section I. GENERAL 11. SCOPE. Chapter 3 contains information for first and second echelon maintenance of The Short Wave Diathermy Apparatus. There are no organizational tools or equipment for use in conjunction with this equipment. Caution: NEVER remove doors of cabinet unless line cable is dis- connected from receptacle. This is a precautionary measure due to high voltage. Section II. LUBRICATION 12. BLOWER MOTOR. The exhaust blower motor must be lubricated every 3 months with SAE No. 30 engine OIL (OE). Remove the four screws holding the blower panel to front of cabinet, as shown in figure 7. With one hand, hold the top of front door and bottom of blower panel, and tilt top of blower panel forward exposing oil holes in motor, which can be oiled with the other hand as shown in figure 6. Section III. PREVENTIVE MAINTENANCE SERVICES 13. DAILY. Adjust4 ‘trimmer” if required. See paragraph 19. 14. THREE MONTHS, a. Lubricate blower motor. See paragraph 12. b. Dust. See paragraph 18. Section IV. TROUBLE SHOOTING 15. WHITE PILOT LIGHT NOT LIT WHEN MAIN SWITCH IS TURNED ON. Possible Cause Line fuse blown. Line cord broken or not plugged in. Pilot bulb burned out. Possible Remedy Replace line fuse. Replace or insert plug. Replace pilot bulb. 16. PLATE CURRENT TOO HIGH. Possible Cause “Trimmer” out of adjust- ment. Possible Remedy Adjust “trimmer”. See para- graph 19. 17. HEAT NOT REACHING PATIENT BUT VOLTMETER REGISTERING Possible Cause Patient cable not plugged in. Possible Remedy Plug in patient cable. Section V. MAINTENANCE OPERATIONS 18. DUSTING- Every 3 months, or oftener if necessary, the dust should be cleaned off of the parts inside of cabinet. The best way to do this is to remove both front and rear doors and use an electric hand blower, suction pressure apparatus, or compressed air. DO NOT wipe with a cloth as this will disturb parts that might be thrown out of adjustment. Be careful not to blow dust under dust cover, 7R08288 (fig. 3), on relay panel. Med. Dept. No. 1 SR00107 Screw, 6-32 XI/2-Inch, R.H.M. Nomenclature Figure 6. Oil holes. 19. “TRIMMER” ADJUSTMENT, a. If any of the following actions occur when the oscillator switch is turned on, the “trimmer” must be adjusted. (1) Plate current exceeds 150 ma. (2) Plate current does not reach a minimum of 90 ma or less, even after operating 10 minutes. (3) Plate current increases instead of decreases. b. “Trimmer” adjustment. With power control and resonance control on ZERO, and with nothing plugged in outlet panel, operate the unit 12 with oscillator switch on for at least 10 minutes. Remove cover plate, 7R08250 (see figs. 2 and 8), and loosen lock screw shown. Move the adjusting lever, 7R08190 (fig. 8), up or down very slightly and watch plate current milliammeter. The purpose of this adjustment is to tune the “trimmer” condenser for minimum plate current, and it has to be done very carefully by moving the lever very slowly. If the lever is moved too far, the plate current may exceed 350 ma and trip the overload relay. The current is restored by turning power control to RESET. Rock the lever, 7R08190 (fig. 8), slowly up and down until the minimum plate current position is found, and then tighten the lock screw, SR00611 (fig. 8), and replace cover plate, 7R08250 (fig. 8). If the plate current exceeds 150 ma, even after operating a few minutes, it is best to make a preliminary adjustment first and then wait 10 minutes for a final adjustment. Med. Dept. No. 1 SR00107 Screw, 6-32 X 1/2-inch, R. H. M. Nomenclature Figure 7. Holes for oiling blower motor. c. If the “trimmer” adjustment has been made as outlined in paragraph 19, and it is impossible to get a minimum plate current of 90 ma or less, cr if the overload relay opens the circuit whenever the oscillator switch is turned on, then it will be necessary to have the tubes and neutralization checked. d. If, sometime after the “trimmer” adjustment has been made, the no load plate current does not decrease to 90 ma or less in 4 minutes after turning on oscillator switch (with power control and resonance control on ZERO), then the adjustment must be repeated. Med. Dept. No. Med. Dept. No. Nomenclature Nomenclature 1.7R08250 Plate, Cover, "Trimmer" Condenser. 2. SR00611 Screw, 6 32 X 3/8-inch, Fill. H. M. 3.7R08190 Lever, Adjustment, “Trimmer” Condenser, Complete. Figure 8. “Trimmer” adjustment. APPENDIX LIST OF ALL SERVICE PARTS O EM NO. 7105505 DIATHERMY APPARATUS, SHORT WAVE, COMPLETE, 110 VOLT, 60 CYCLE Burdick Corporation Model No. SWDX-80 Fig. No. Medical Dept. No. Nomenclature Unit Quan- tity SR 00003 Common Parts Screw, 8-32 X R. H. M., 144topkg. Used pkg- — SR00010 in resistor bank frame. Screw, 8-32 X %-inch, R. H. M., 144 to pkg, Used Pkg. * SR00015 in various mountings. Screw, 8-32 X 1-inch, R. H. M., 144 to pkg. Used pkg. SR00017 to secure FP 265 tube socket. Nut, 8 X 32, hex, 144 to pkg. Used in various Pkg. SR00040 mountings. Screw, 6-32 X %-inch, R. H. M., 144 to pkg. pkg- * SR00042 Used to secure bottom screen. Washer, screw size 6, 1 lb. pkg., 1300 washers. Pkg- * SR00043 Used in various mountings. Nut, 6 X 32, hex, 144 to pkg. Used in various Pkg- *4 SR00061 mountings. Tube, amplifier, beam tetrode, GL—814 ca. 1 *4— SR00062 Tube, oscillator, beam tetrode, metal, 6L6 ea. 1 *4 SR00063 Tube, rectifier, mercury, 866—A ea. 2 *4 SR00064 Tube, rectifier, 5Y3-GT ea. 1 *_ SR00065 Resistor, 100-watt, 3000-ohms, wire-wound, vitre- ca. 1 *— SR00066 ous enameled. Resistor, 25Twatt, 5000-ohms, wire-wound, vitre- ea. 1 * SR00067 eous enameled. Resistor, 25-watt, 7000-ohms, wire-wound, vitre- ea. 1 * SR00068 ous enameled. Resistor, 25-watt, 1500-ohms, wire-wound, vitre- ea. 1 *_ SR00069 ous enameled. Resistor, 100-watt, 35000-ohms, wire-wound, vitre- ea. 1 * SR00070 ous enameled. Resistor, 100-watt, 25000-ohms, wire-wound, vitre- ea. 1 *5 SR00071 ous enameled Bulb, pilot, 6—8 V., miniature screw base ca. 2 *— SR00105 Screw 6-32 X %-mch, R. H. M., 144 to pkg. pkg. * SR00106 Used in various mountings. Screw, 6—32 X R. H. M., 144 to pkg. Pkg- *6-A SR00107 Used in various mountings. Screw, 6-32 X PJ-inch, R. H. M., 144 to pkg. Used Pkg- ♦ SR00108 in various mountings. Screw, 6-32 X %-inch, R. H. M., 144 to pkg. Used Pkg- See f 1 in relay board and resistor mountings, ootnote at end of table. Fig. No. Medical Dept. No. Nomenclature Unit Quan- tity — SR00115 Screw, 8-32 X %-inch, R. H. M., 144 to pkg. Used in frame clamps. Pkg- — SR00151 Washer, lock, screw size 10, 1,000 to pkg. Used in blower mounting. Pkg- * SR00152 Washer, lock, screw size 8, 1,000 to pkg. Used in various mountings. pkg. ♦ SR00153 Washer, lock, screw size, 6, 1,000 to pkg. Used in various mountings. Pkg- * SR00154 Washer, lock, screw size 4, 1,000 to pkg. Used in meter mounting and power control. Pkg- — SR00156 Washer, lock, screw size 5/iq, 1,000 to pkg. Used in various mountings. Pkg- ' SR00226 Screw, %-20 X % inch, R. H. M., 144 to pkg. Used in various mountings. Pkg- — SR00229 Washer, lock, screw size %, 1,000 to pkg. Used to secure autotransformer. pkg. — SR00230 Nut, % X 20, hex, 144 to pkg. Used to secure auto- transformer. Pkg- * SR00244 Washer, screw size 4, 1 lb. pkg., 2,400 washers. Used in meter mounting and power control. Pkg- * SR00255 Screw, 6-32 X inch, Phillips O. H. M., 144 to pkg. Used in treatment drum. Pkg- * SR00265 Screw, 8-32 X %-inch, R. H. M., 144 to pkg. Used in various mountings. Pkg- SR00293 Washer, screw size %, 1 lb. pkg., 144 washers. Used in autotransformer mounting. Pkg- * SR00304 Washer, screw size 8, 1 lb. pkg., 1,300 washers. Used in various mountings. Pkg- — SR00316 Plug, ground, male ea. 1 — SR00317 Plug, ground, female ea. 1 — SR00324 Screw, 8—32 X %-inch, FI. H. M., 144 to pkg. Used in top cabinet frame. Pkg- * SR00373 Setscrew, 10-32 X %-inch, Bristo head, cup pt., 100 to pkg. Used in plastic control knobs. pkg. *— SR00377 Screw, 4-40 X inch, R. H. M., 144 to pkg. Used in meter mountings. Pkg- * SR00385 Nut, 4 X 40, hex, 144 to pkg. Used in meter mount- ing and power control. Pkg- *— SR00391 Setscrew, %-20 X %-inch, Allen head, cup pt., 144 to pkg. Used in treatment drum. Pkg- * SR00414 Screw, 4-40 X %-inch, R. FI. M., 144 to pkg. Used to secure cable on control pulley. Pkg- — SR00415 Washer, screw size 8, fiber, 500 to pkg. Used in FP 265 tube socket mounting. Pkg- SR00416 Screw, 10-24 X 1%-inch, R. H. M., 144 to pkg. Used in blower mounting. pkg. SR00417 Nut, 10 X 24, hex, 144 to pkg. Used in blower mounting. Pkg- SR00418 Screw, 4-40 X 6-inch, O. H. M., 144 to pkg. Used in shaft guide. Pkg- SR00419 Screw, 7 X J4-inch, sheet metal, B. H., 144 to pkg. Used in front inner mounting panel. Pkg- SR00420 Screw, 6-32 X 1%-inch, R. H. M., 144 to pkg. Used in 866—A tube socket mounting board. Pkg- SR00421 Screw, Yie-18 X 1%-inch, hex H. M., 144 to pkg. Used to secure piste transformer. Pkg- See f( SR00422 jotnote at en Nut, X 18, hex, 144 to pkg. Used in plate trans- former mounting. d of table. Pkg- Fig. No. Medical Dept. No. Nomenclature Unit Quan- tity — SR00423 Cord, type “S”, hard service rubber, No. 14, three- conductor. ft. io y2 — SR00424 Plug, male, two-prong, rubber covered ea. i — SR00425 Plug, female, twist lock, three-wire, Hubbell No. BC-11357. ea. i 2 SR00426 Receptacle, male, twist lock, three-wire, Hubbell No. 7556. ea. i ■ - SR00427 Bracket, resistor, ohmite No. 12 ea. 12 — SR00428 Bracket, resistor, ohmite No. 26 ea. 2 — SR00429 Washer, screw size 6, fiber, 500 to pkg. Used in resistor mounting. Pkg- SR00430 Screw, %6-18 X %-inch, hex head machine, 144 to pkg. Used to secure base to cabinet. Pkg- — SR00437 Screw, 10 X 1%-inch, R. H. Wood, 144 to pkg. Used to secure cable rack. Pkg- ♦ SR00538 Screw, 6—32 X %-inch, FI. H. M., 144 to pkg. Used to secure main switch. Pkg- *2 SR00575 Screw, 6-32 X FI. H. M., 144 to pkg. Used in rear panel and top. Pkg- * _ SR00576 Screw, 4-40 X %6-inch, FI. H. M., 144 to pkg. Used to secure oscillator switch. Pkg- 7 SR00611 Screw, 6-32 X %-inch, Fill. H. M., 144 to pkg. Used on “Trimmer” condenser. Uncommon Parts Pkg. *4 7R08102 Crystal, quartz ea. 1 *3 7R08104 Tube, amplifier, high mu triode ea. 2 * 7R08106 Resistor, 2-watt, 100,000-ohm Resistor, 2-watt, 400-ohm ea. 1 ♦ 7R08108 ea. 1 * 7R08110 Capacitor, mica, 50-mf., 500-v 1 ♦ 7R08112 Capacitor, mica, .005 mf, 500-v ea. 3 ♦ 7R08114 Capacitor, mica, .05 mf, 500-v ea. 1 ♦ 7R08116 Capacitor, mica, .03 mf, 1000-v ea. * 7R08118 Capacitor, mica, .005 mf, 2000-v ea. 4 ♦ 7R08120 Capacitor, mica, .002 mf, 3000-v ea. 1 * 7R08122 Capacitor, paper, .1 mf, 200-v ea. 2 * 7R08124 Capacitor, paper, .1-1 mf, 200-v ea. 2 ♦ 7R08126 Capacitor, electrolytic, 8-8-8-mf, 450-v ea. 1 ♦ 7R08128 Capacitor, tubular, .1 mf, 600-v ea. 2 * 5 7R08130 Voltmeter, triplett ea. 1 * 5 7R081.32 Milliammeter, triplett ea.. 1 * 7R08134 Choke, iron core, Thordarson ea. 1 * 7R08136 Choke, radio frequency, RFC-1 ea. 1 ♦ 7R08138 Choke, radio frequency, RFC-2 ea. 1 ♦ 7R08140 Choke, radio frequency, RFC-3A-3B ea. 1 * 7R08142 Relay, complete: assembly ea. 1 ♦ 7R08144 Breaker, circuit, main switch ea. 1 ♦ 7R08146 Cable, treatment ea. 1 * 2 7R08148 Drum, treatment, complete. With cable less arm. . ea. 1 ♦ 7R08150 Spacer, cable ea. 1 ♦ 7R08152 Sleeve, cable, rubber ea. 1 ♦ 7R08154 Cable, coupling coil pulley ea. 1 * 7R08156 ea. 1 * 7R08158 Tool, bulb, neon ea. 0 * 7R08160 Tool, screw driver, short ea. 0 ♦ 7R08162 Tool, screw driver, long ea. 0 * See f 7R08164 ootnote at en Tool, clip and bulb d of table. ea. 0 Fig. No. Medical Dept. No. Nomenclature Unit Quan- tity *5 7R08166 Switch, toggle, oscillator 1 1 ♦ 7R08168 Cover, treatment drum, plastic ca. *1 7R08170 Screw, %-20 X % inch, thumb, brass. Used to secure front panel. ea. 2 ♦ 7R08172 Knob, round, plastic. Used for various controls. . ca. 3 * 7R08174 Spacer, metal. Used in patient pull off lever ea. 1 *5 7R08176 Lever, patient pull off, complete. With ball and cord. ea. 1 + ♦ 7R08178 7R08180 Socket, screw, pilot bulb, complete. Assembly. . . . Insulator, mounting, hole, porcelain, com- plete. Assembly. ea ea. 2 ♦ 7R08182 Insulator, %-inch hole, rubber ca. 2 ♦ 7R08184 Board, terminal, autotransformer, complete. As- sembly. ea. 1 ♦ 7R08186 Insulator, %-inch hole, rubber ca. 1 *3 7R08188 Clip, plate, amplifier tube, complete. Assembly. . Lever, adjustment, “Trimmer” condenser, com- plete. Assembly. Lock, “Trimmer” condenser adjustment lever, complete. Assembly. eg. 2 *7 7R08190 ea. 1 * 7R08192 ea. 1 * 7R08194 Board, terminal, relay, complete assembly ea. 1 — 7R08196 Resistor, 100-watt, 3,000-ohms ea. 2 — 7R08198 Transformer, plate high voltage power, supply. . . . ea. 1 — 7R08200 Autotransformer, 100-130-v., 50/60 cycles ea. 1 — 7R08202 Transformer, plate, low voltage power supply ca. 1 5 7R08204 Switch, voltage control, complete. Assembly ea. 1 — 7R08206 Socket, FP—265 tube ea. 2 — 7R08208 Socket, GL—814 tube ea. 1 — 7R08210 Socket, 6L6 tube ea. 1 — 7R08212 Socket, 5Y3-GT tube ea. 1 — 7R08214 Socket, 866—A tube ea. 2 — 7R08216 Socket, crystal ca. 1 — 7R08218 Blower, complete. Assembly ea. 1 — 7R08220 Panel, blower mounting, complete. Assembly. . . ca. 1 — 7R08222 insulator, felt, 1-inch wide, thick, 9 inches long. For blower mounting. ea. 1 — 7R08224 Insulator, felt, long, 1-inch wide, %-inch thick. For blower mounting. ea. 1 — 7R08226 Ventilator, blower, metal ea. 1 — 7R08228 Guide, shaft, control. Used on control shaft assemblies. ea. 2 — 7R08230 Strap, ground, woven, copper. 6 inches long.... Wire, glass insulated tips, woven copper cover, complete. With lamps. ca. 1 7R08232 ea. 1 — 7R08234 Panel, patient electrode receptacle mounting, com- plete. Assembly. ea. 1 — 7R08236 Insulator, patient electrode receptacle, plastic ca. 2 7R08238 Receptacle, patient electrode, complete. Assem- bly. ea. 2 — 7R08240 Frame, resistor bank mounting ea. 1 — 7R08242 Board, asbestos, resistor mounting ea. 2 7R08244 Board, mounting FP-265 tube socket, complete. Assembly. ea. 1 — , 7R08246 Panel, exciter power supply mounting, complete. . . ca. 1 — 7R08248 Connector, plastic. For blower motor leads Plate, cover, “Trimmer” condenser ea. 2 2-7 7R08250 ca. 1 2-3 See f 7R08252 sotnote at en Holder, treatment arm, complete. Assembly d of table. ea. 1 Fig. No. Medical Dept. No. Nomenclature Unit Quan- tity 7R08254 Panel, instrument, mounting-, complete. Assem- ea. 1 — 7R08256 bly. — 7R08258 Frame, top, cabinet ea. 1 Frame, tubular, metal ea. 2 — 7R08260 Clamp, tubular frame ea. 16 — 7R08262 Cabinet, metal, complete. Assembly ea. 1 — 7R08264 Base, metal, complete. Assembly ea. 1 — 7R08266 Panel, front, inner mounting ea. 1 — 7R08268 Cover, front, inner mounting panel ea. 1 — 7R08270 Panel, rear, inner mounting . ea. 1 — 7R08272 Screen, wire, bottom ea. 1 — 7R08274 Caster, swivel, 3-inch, complete. Assembly ea. 4 - — 7R08276 Rack, treatment cable ea. 1 2 7R08278 Arm, treatment drum, adjustable ea. 1 3 7R08280 Plate, shield ea. 2 5 7R08282 Adjuster, voltage ea. T 5 7R08284 Control, resonance ea. 1 5 7R08286 Control, power ca. 1 3 7R08288 Cover, relay ea. 1 *— To be requisitioned, when required, from the supply depot. No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition. ☆ U. S. GOVERNMENT PRINTING OFFICE: 1945—6S7664 TM2003