TECHNIQUE 0 F MARROW NAILING by GERHARD KUENTSCHER 'rof. Dr, med. habil# RICHARD MAATZ Dozent Dr, med, habil. Assistants of the Surgical Clinic at the University of Kiel With 101 Illustrations 19 4 5 GEORG TRIE ME EDITION LFIPZtG Translation prepared by: Office of Military Government for Germany (U. S.) Office of Naval Advisor Medical Section Foreword to translation of TFCHNIQUE OF MARROW NAILING by KUENTSCHER - MAATZ This translation is included here for historic reasons al- most as much as for medical reasons. It appears to be the first b»»k about the Marknagel, Although KUENTSCHER is listed as senior author there is good reason to believe that he had little to do with the writing of this book* MAATZ, who it appears was the author was junior to KUFNTSCHJ at Kiel. MAATZ is said to have been educated as an engineer before he studied medicine. The multiplicity of ideas and modi- fications of the basic concept of the nail have been cited as evidence of his engineering background. Certainly in the United States he would be spoken of as a"gadgeteer", Without doubt he is a capable surgeon and qualified to have his opinions. The personality of two people could not be more opposite in nature. MAATZ the tense, extrovert rapid fire react- ions - KUrNT£CHER the shy scholarly introvert, self deprecating and carefully thinking things through. One canrot escape the thought that the Surgical Clinic at the University of Kiel must have been an interesting place when these two men were debating a subject. The publication of this booklet was delayed from 1942 until in 1944 because of a government restriction on paper. In late 1944 an edition of 1000 copies was authorized. approximately half of these had been distributed at the end of the war. The fate of the other copies and the plates in Leipzig is not known. There has been no practical way of reproducing the illus- trations in the original text* Since many explained by line drawings these have been traced. Some have been omitted entirely as unnecessary or not suitable for reproduction with the facilities at hand, the important details being lost in the tracing on a mimeograph stencil* Upon the completion of this task the original text with any reserve copies of the translation will be sent to the Department •f the Navy, Bureau of Medicine and Surgery, Publications Divi- sion. HARRY J. ALVIS Commander MC U. S. Navy PREFACE The idea that a fracture may be held in good position un- til healed by introducing foreign bodies into the marrow cavity is not new, Metal wires and pins and even ivory bolts have been used for this purpose. Without an exposure of the broken ends and the associated danger of infection the introduction of the above mentioned foreign bodies could not be accomplished. The essential feature of KUFNTSCHFR*s method is the intro- duction of a foreign body from a place apart from the fracture site. In this it differs essentially from all other methods of operative treatment of compound fractures. Through the use of a specially shaped, very stable, properly constructed, nail- like splint, which forges an elastic union with the inner surface of the marrow cavity, the greatest possible stability of the fracture should be achieved, so that a further support by plaster cast or traction apparatus is not required. By this the disad- vantage of a prolonged immobility of the limbs is avoided, KUFNTSCHER has extended the principle of the nailing of the neck of the femur (SMITH-PETERSEN) to the shaft of bones. That KUENTSCHER«s method was at first generally disapproved of when he announced it at the Surgical Congress in Berlin, is understandable. This disapproval only confirms the conscientious- ness of the German Surgeons, Above all they were concerned that the introduction of such a large foreign body into the marrow cavity of a bone might be harmful to the patient. I hid the same apprehension myself, when KUT’NTSCHER expounded his idea to me for the first time even aft°r the results of animal experi- ments were known. This "instinctive" apprehension was not sub- stantiated, Subseouently no undesirable result was observed in humans in the marrow cavity or elsewhere. The method has now been adopted in numerous clinics and hospitals. It is now being constantly developed. In different places, not only with us at the clinic in Kiel, surgeons try to perfect the construction of the instruments, improve the pro- tection of the doctor from X-rays, and the form of the nail. Besides all these technical matters th<"re is constantly the ouestion of the indications. Gradually it can be ascertained whjch fractures are suitable, which are less suitable and which are entirely unsuitable for nailing; for which type of fracture in spite of the nail, a cast is still required, which fracture can be subjected to weight-bearing early and which one late. All these matters are still under investigation. To all surgeons now engaged in the nailing operation or to those who intend to practice it on this treatise should be a help. It was written to make them acquainted with the present state of our knowledge and besides this to point out the threatening mistakes. I would like to point out that only the most skilled surgeons acquainted with the treatment of fractures, should be allowed to "nail" and even then only when having a complete armamentarium at their disposal. The nailing operation requires the greatest technical skill. It is my opinion that KUENTSHCER*p method is the greatest advance in the handling of fractures given to ihe world since the description of the extension nail method of KLAPP, I am quite sure that its use will spread throughout the world, ii is not difficult to bid this book Good'-speed, as I am coovinCed that it will make its own way* A, Wi Fischer, Kiel PREFACE In December 1942 we had completed for the first time the following description of the marrow nail operation. However, permission to publish the treatise was not given. When these difficulties had been overcome after the lapse of a year, the entire material of illustrations was destroyed during a severe air raid on Leipzig. By my new assignment to sea duty, contact with my publishers and my collaborator has become very difficult. For these reasons the publication of this work was repeatedly de- layed. Unfortunately we are also not in a position to print an annex to this treatise with a short summary of the results we have had in the meantime, for, more than 500 nailing operations have been performed since this work was written of which about 400 have been done at the Kiel clinic and the rest in the field or in station hospitals. In the meantime all of us have learned to resort to makeshift, much more. It is not possible at present to get all the instru- ments described in this book because their construction had to be stopped due to circumstances of the war, Nevertheless it is possible to perform nailing operations with all the respons- ibility required if the surgeon has at his disposal X-ray appar- atus, nails, awls, hammers and block-and-pulley. All the other instruments needed can be replaced and are not absolutely re- quired. Every beginner is urgently recommended to use at first only the strict indications and to start only with nailing of the most suitable fractures. In this case he will find full satis- faction with this wfnderful method. In the course of time the indications will extend themselves. In every case one must bear in mind the necessity of accomplishing a stable union of the ends of the bones by nailing. If this condition is not fulfilled fail- ures will not be the fault of the method but of the operator. On board a hospital ship in the Mediterranean August 1944 M a a t z INDEX PREFACE 2 PREFACE 4 1* Introduction 7 2 . Tools 15 3. The indication and detailed statements about the after- treatment 23 A. The thigh 25 B . The Leg 27 C. The upper Arm ., 34 D. Forearm 36 E. Collarbone 37 4. Suitable time and preliminary treatment , 37 5. Narcosis 3B 6. Appropriate position and adjustment 39 A. The thigh 39 B. The Leg ,. 40 C. The upper Arm 41 D. Forearm 41 7. Fluoroscopy 42 8. Asepsis 44 9. Selection of the Nail 44 Thickness 44 Thigh 45 Leg . 45 Upper Arm 45 Forearm 45 Length , 45 Thigh ,, 46 Leg 46 Upper Arm . . 46 Forearm 4 6 The longitudinal profile 46 10, Insertion of the guide rod 47 Thigh 43 Ulna 49 11, Insertion of the Nail 49 12, The After-Treatment 53 13, Removal of the Nail 54 Suitable time 54 Technique 55 Thigh .. 56 leg 56 Upper Arm , . 5£ Forearm 56 14* Nailing in the case of an open fracture 57 15, Nailing of gunshot fractures 59 16, Nailing with delayed formation of callus 60 17, Nailing operation in pseudarthroses 61 18, Osteotomy 61 19* Nailing of a spontaneous fracture ,, 20. The Y-Nail * . * . . 4 > 21* The supracondylar T-fracture 75 21, Atypical nailings 76 23* Bibliography ...... . r3 7 1*_introducti£N Before the surgeon undertakes his first marrow nail oper- ation he must have a clear understanding of its main principles. Only then will he be in a position to apply the method as widely as the present state of the technioue makes it possible. It must not be forgotten that it is a case of true nailing. Just like the carpenter's nail unites two pieces of wood with elastic strength, so the marrow nail is meant to unite the bone ends. The nail should be introduced in such a way by strokes with a hammer into both portions of the broken shaft of the bone that it sticks there and makes it impocBlble to shift either end of the broken bone shaft. Only by accomplishing this can one speak of an entirely satisfactory marrow nail operation and only then can its great advantages over all other methods be turned to account, A further securing of the fracture by splints, plaster cast or extension will the,n not be required. The limb may be freely moved, even bear weight. Wasting of the muscles and disturbances of the circulation will not occur, or, if so, only to an unim- portant degree. Joint stiffening is not to be feared for the joint is kept at rest. The fracture, placed under favorable mechanical conditions will be bridged over by bone in a very short time, for it is subjected only to pressure while pulling and pushing tensions are avoided. The nail is held by the friction of the rough inner layer of the cavity of the hollow bone cylinder.. Since the inner diameter of the bone varies it is of the utm.ost importance that the nail should have as much elasticity as possible in the cross section. Therefore the nail is U-shaped in the cross section. After a successful nail operation the friction bet- ween the- bone and the nail may be s« great that the ends of the bone are held together by a force whi• h can a pull of several hundred kilograms. But not all broken bones may be treated under equally favorable conditions. The shape of the bone, the shape and location of the fracture and finally the possibilities of introducing the nail into the cylinder of the bone are all of significance, A thorough knowledge of the forces in ouestion in the individual case, both for the indications and for the technique of nailing and particularly for the treat- ment after the operation is absolutely necessary. It will not be possible to achieve the aim of the marrow nailing in every case, especially sufficient friction of the nail in both ends of the broken bone. But nevertheless, surgeons may use this method and successfully apply it. It is only important to know how to avoid the wpak points of this method. The mechanical factors are simple and easy to understand, A bone with only insignificant variations of the diameter of the cavity presents the m©st favorable conditions possible, (in. i). Illustr, 1 The location and the shape of the fracture are of no import- ance if the two fragments are sufficiently long to give the nail a satisfactory hold. In the case of a comminuted .fract- ure, however, the nail alone has to withstand all the forces tending to produce angulation at the fracture site. While in the case of a plain transverse fracture the nail is held by the bone in a firm mechanical system so that the nail alone does not have to withstand the forces tending to produce angulation at the fracture site. When the friction between the nail and the bone is in sufficient, the nail will act as bolt (ill, 2), Ill uvt r, 2. 9 Indeed, it prevents a longitudinal bending and lateral displace- ment, but permits rotation, In this case shape of the fract- ure is of decisive importance, A comminuted fracture will short- en and a spiral fracture will be distorted. Fven true transver- se fractures would not be protected against rotation. On the other hand, oblioue fractures usually shorten only as long as there are minor faults of rotation, until finally the nail (bolt) and bone Jam, (ill, 3) In the case of a gagged fracture the 111 u st, r. 3 bone ends can, nevertheless, be united in such a way that the bone ends remain firmly fixed together (so-called Hirth-Teeth). The muscles press the bone ends of the fracture together and in this way rotation is prevented. The outlook is not so favorable in the case of a bone with*! considerable variation in the diameter of th° marrow cavity. If the fracture lies in the narrowest part of th° marrow cav ity , a successful nailing can still be achieved uwing to the elastici- ty of the cross section of the nail (ill, 4). For the distance Titudr.H of several centimeters the nail has sufficient friction in the marrow cavity of both fragments. Should this friction be insig- nificant the nail will act as a b»lt and will prevent only a lateral displacement, In the case of a transverse fracture the bone ends, which are in good position and ouite secure against lateral displacement will be firmly held togeth-r by the muscles and angulation cannot occur, and rotation movements are prevent- ed by the interlocking of the jagged If a of bone is missing at the fracture site the bone ends cannot stand one upon the other longtudihally and will be a tendency to angulation (ill. 5j. If the fracture does not lie in the narow- a Thigh 111. 5 i. % • % Ug est part of the marrow cavity the degree of disproportion bet- ween the width of the marrow cavity and the thickness of the nail will determine whether the displacement c n be overcome by the nail or not. If 'the broken ends can be displaced laterally by the thickness of the bone cortex we then find the most un- favorable conditions. They are: lateral dislocation, distraction, angulation and rotation i.e.. which are all the faulty positions that can be imagined (ill. 6). 111. 6 Unfavorable conditions for marrow nailing The situation is ior° complicated when the nail is intro- duced into the bone shaft from the side. In the fragment into which the nail has been introduced from the side a rotation is impossible between the nail and the bone shaft but in the other fragment the much desired strong friction uniting nail and bone will generally not be achieved, because the nail must be somewhat smaller than the marrow cavity (ill, 7), When in this case the Illustr. 7 a - favorable b - unfavorable fracture is situated at the narrower part of a tap°ring marrow cavity, the difference of the respective calibers, which is neces sary f©r the introduction »f the nail, is to no disadvantage. The greater elasticity of the nail, which is ary for its lateral introduction in these cases is an undesirable feature. As a further place of friction between the hail and the hen'.; we have the spongiosa of the ends of the bones. While in the majority of cases the stability ©f the nail in these marts of the bone is irrelevant to the stabilization of a fracture, it is nevertheless of special significance in sov cases, and one must understand clearly how much pressure this soft bone tissue can withstand. For example, one may mention the transplantation of th» radius to the ulna in the vicinity of the elbow joint, A bony healing was not achieved after five months (infection preventing the effect of the nail,). The joint was practically immobile. The motion that was possible from 100° to 80°, was not due to the elasticity of the nail, but as the pictures show, the nail had worked a slot in the spongiosa of the bone, in which the end of the nail covered a distance of 12 millimeters (ill,8), Illustr, 8 The b«ne tissue subjected to excessive pressure rhythmical1y or constantly reacts by absorption at the point of pressure. While in the tubular part, with pressure of the nail fror* the inside and absorption of the bone at the inner wall, the peri- osteum prevents a lateral breaking through of the nail by a speedy formation of new bone. Such a reaction is not s°en in spongy bones. The nail works itself a slot and finally breaks out »f the bone (ill,9), Therefore a subtrochanteric fracture Illustr, 9 must not bear weight too early, and the head of the nail must be particularly broad for fractures the elbow joint, and, for the same reason, it is useless to insert a nail in a fracture of the tibia so far that the point of the nail gets a hold, in the distal spongy portion of the bone. This hold is negligible and unreliable, •ur aim and purpose must be the nailing operation. It is not always to be achieved. At least the principle of the bolt should be guaranteed. It still offers great advantages. If a baiting cannot be achieved with certainty the fracture is un- suitable for the new method. Of course, even hr re there are exceptions to be made and in certain cases one will decide to nail the more readily when the fracture can be brought to a good healing only with difficulty by the methods used previously. As examples one may refer to the comminuted fractures shown in the pictures above, which were nailed with the very best results, although the mechanical conditions were by na means ideal ones <111, 10 and 11), In practice the conditions in sore casns may be very com- plicated , So, i or instance, rhe analysis of the forces whi ch beet.' a tibia fracture in good position with a spread nail is very diffi- cult and it is impossible to decide wheth r the security of the fractures is due mostly to the principle of the nail, the bolt, the Hirth teeth or other factors. The o n 13r real important thin a is that the Cioctor should attain sufficient skill to r °co? nize a, s soon as possible to which extent a fracture, tr ated in this way may be subjected to functional use. For this, it is only poss- ible to give directions. Personal experience alone en- sures a masie / of this subject. The foregoing statements make it clear that one cannot speak simply of a.n indication for using the marrow nail. It is necessary to distinguish between "strict" and ‘extended'' indi- cations. In the first group may be included t-ose fractures Illustr. 10 Illustr. 11 which in the following treatise are considered "particularism suitable", i.e, those fractures which can be treated by the nailing operation much better than by any method. In the other group are included the fractures considered "suitable" and "less suitable" and at the same time those which do not show all the advantages of the new method aft°r they are nailed. For the beginner It is advisable to undertake only those cases which fulfil the "strict" indications. kith growing ex- perience operations may be performed on cases which are classi- fied under the "extended" indications. While the nail is In place certain changes occur which must be taken into consideration. The nail may become loose in the bone as early as the first week. The inner wall of the bone is then worn thinner by the pressure of the nail. The conseeuence is that the friction between the bone and the nail becomes less. As absorption of the bone occurs only at the place of the strong- est pressure, a groove in the marrow cavity develops, which per- mits motion of the fragments in the axis of the bone. Rotation- al displacement i s, ho¥fever, prevented by the callus which fills the broken ends. Thus there exists the most favorable conditions for the healing of the bone, as the nail acts as a sliding splint and permits only the forces pressing the ends together to act. It stands to reason that the locsening of the nail after a couple of weeks may be a great disadvantage, according to the form and position of the fracture, therefore it must be con- sidered accordingly, A blocking effect of the nail is irEelevant if the nail provides sufficient stability and has a sufficient hold for the fragments. The bone gap will be bridged over in a surprisingly short time,. ... . If a sufficient stability of the nail or sufficient frict- ion between the nail and the bone does not exist the blocking, effect of the nail may result in a delayed healing of the fracture, The point of the nail, acting as a bolt, produces a con- stantly widening slot in the short fragment, and thus a later- al piercing by the nail of the periosteal callus which is under pressure is prevented. The fragments are not fixed firmly. How strong the blocking effect of the nail may be is proved by the fact that nails, under the above mentioned conditions, frequent- ly begin to move and thus work their way out of bone* The moving of the nail is always a proof of an insufficient stabilization of the fracture by the nail, A fracture is never firm as long as it allows the nail to work itself out. Even an apparently well-fixed nail in a fracture favorable for nailing may lead to disturbances of the healing process a:d finally to considerable delay in th*5 format! n of callus, if even a single mistake is made. If in a smooth transverse fract- ure in the middle of the leg, for instance, a nail is introduced that is too thin, this may permit shearing forces to act at the fracture site, which does not permit a bony healing at al" or only a very slow healing. The reason is that gradually the nail will bend a little, and at the same time the upper part of the fracture will move for?/ard and thb lower part backward as much as the difference between the diameter of the nail and the marrow cavity permits. When the weight is taken off the elastic for- ces of the nail will push the bone ends back to their former position. Even if the distance is only J or 1 millimeter this displacement will be enough to delay a bony healing or even to make it impossible (ill, 12). Illustr. 12 Here are some other brief comments important for consider- ations in connection with the nailing. As the fracture is nailed from a considerable distance and the operation wound will be only some centimeters long at the most the danger of an infection will be very slight. The nailing is not a serious operation which might endanger the life of the patient. Just like operations, the nail- ing too must, of course, not be done in shock. An insignificant fat embolism is to be expected (MAATZ), Clinically it does not appear but even this slight additional hazard must not be added to cases already in a precarious state, A good formation of callus may be expected only if a stabile osteosynthesis has been attained. A very abundant formation of callus may be caused by the issue of the marrow at the fracture site (KUFNTSCHER). Nothing has said yet about the fixing of the fracture. The chemical stimulation by the metal of the nail may cause a far reaching disorganization in the nailed bone and along with this an abundant periosteal formation of callus. This occurrence is to be considered as undesirable. The fixing of the fracture may even be prevented (MAATZ). Therefore th15 ncf. should be made of corrosion-pro$f metal. The most careful consideration of the following technical directions is urgently recommended. The nailing operation is a kind of operation which owing to its technical character is not in the line of every surgeon. There are easy as well as difficul cases of nailing and one should always be prepared for the lattei Never make the mistake of considering a case of a nailing oper- ation afinfinished and dene with”. For a swccec.aful nailingoperataton a ways requires a careful clinical and ambulatory observation to meet every complication wh'-ch may occur in the course of time, only can the new method show how vastly superior it is in comparison with the nitherco used methods, TOOLS N.l'LS The above described mechanical conditions of the marrow nailing operation make it clear that the marrow nails now avail- able do not make it possible to obtain a stabile osteosynthesis in every case. The nail lies close to the inner wall of the bone shaft not only at the fracture site but also beyond it as for as possible. To attain this end the nail must have great elasticity in the cross section because the diameter of the marrow cavity varies considerably. This required elasticity in the cross section is much greater than that attainable by the most favorable cross-'section design of the nail if the thick- ners of the metal is adeouate, Therefore nails must be cr structed which can be spread at the desired spots and, further d.. v' which make it possible to furnish the nails with cams at the re- quired spots. cams are meant to entr into the inner wall of the bone in order to prevent a rotation of the fragments. When constructing nails of new shapes one must consider that any complicated construction is of no value and must therefore be rejected. One must.nev°r forget that the nail must be strong enough to bear heavy mechanical pressure, breaking of a nail should be an extremely rare occurrence and that the nailing of the marrow should not be reserved for some few highly skilled specialists, but should be available for the use of a majority of surgeons. The circumstances of war with the scarcity of labor and th difficulties of getting the necessary material have of course de- layed any further development so that we hav~ to reckon only wit' the instruments presently at our disposal and so have to get al«ng with some disadvantages. Unquestionably we are at the be- ginning of a new development. Under the prevailing conditions it will be neither possible nor advisable that every hospital or clinic adopting the method of the marrow nail operation will have supply of complete sets of nails. The most commonly used sizes must suffice for th° present and the nails needed for special cases must be ordered by cable. The postponement of the operation caused by this circumstance is only a matter of a short time and there is no euestion that it is justifiable. Nails for the thigh, The cross section is approximately V-shaped. The thickness of the metal sheet, reeuired for the stability of the nail, is somewhat at its back so that the two sides of the nail may easily approach each other (ill, 13). In this way the nail pro- Illustr. 13 vided for particularly wide marrow cavities (to 16 millimeters) may be constructed of a thinner m'tal. Accor "ingly the elastic- - ty of the cross section is increased considerably. Customary sizes are: Adults : Lengths: 30, 34, 36, 40 and 42 centime- ters. Cross Section: 8, 9 and 10 millimeters (c.oncial sizes of 11 - 16 centimeters are often used). Children: Lengths: 24 - 32 centimeters (at inter- vals of 2 centimeters). Cross Section: 6 and 7 millimeters. For the subtrochanteric fracture a conical nail is provided corresponding to the form of the marrow cavity to prevent a la- teral displacement with transverse fractures, and to prevent an inevitable shortening when a nail that is too thin is used (111. l/)_ Illustr. 14 If a bone is to be nailed in which callus is to be expected in the marrow cavity (f.i. refracture or a delayed formation of callus) and if one does not wish to lay open the bone, one may make use of a nail which has serrated teeth at its point (111,15). Illustr. 15 Nail with serrated teeth By means of those teeth the nail can work a bed into the bone for itself, b'ith them the introduction of the nail may be im- possible, (Fditors Note: In such a case, where callus fills the marrow cavity, it is customary to perform the "open" nail- ing, exposing the fracture site and reopening the narrow cavity with a drill or an awl). The Y-nail is a conical thigh marrow nail of 32 millimeters length and a V-shaped cross section of 10 millimeters* A nail which in its cross section is shaped like a T and stands at an angle of 45 degrees to the axis of the long’ nail is introduced through a slot in the marrow nail into the head and the nee1" of the femur. The middle of this slot is about 8 centimeters be- neath the end of the nail head. The cross nail (i-shaped) is 11,5 centimeters long and its he d stands out of the nail about 2 centimeters in order to find sufficient hold in the cortical!s drill hole (ill, 16)* Illustr* 16 For the pertochanteric fractures of -groups III and V the marrow nail must be constructed according to the measurements of the bone because it must lie in the thigh bone so snug that it cannot turn round. The directing_apparatus for the Y-nail, which permits a percutaneous introduction of this nail, will be described in chapter 19, or_t ho_ Nails for the leg ar® double nails. Their cross section *s U-shaped. They are curved, the radius being shorter at the ends, and in the middle almost oo (111, 17) Illustr, 17 Adults : Lengths: 24 - 39 centimeters at intervals of centimeter. Greatest diameter 8 and 9 millimeter^. Children: Lengths: 15 - 27 centimeters at intervals of 1-J- centimeters. Greatest diameter 7 millimeters. With spread^nails, two different shapes been used with success: the double nail with an inclined plane and the turn- spread nail. The former one is a regular double nail for the leg which can be spread in its distal part by an inclined plane and the turn-spread nail. The former one is a regular double nai.n for the leg which* can be spread in its distal part by an in- olined plane attached to th© outer nail (ill, 18). The inclined Illustr. 18 plane (a) and the back wall of the bone tube (b> form the la- teral supports which bond off the inner nail backwards. The location of the inclined plan7* is of the utmost importance for the effectiveness of the nail and must therefore be determined very accurately after studying X-ray, A spread nail having the width •f 8 millimeters and the length of 32 centimeters the inclined plane of which is 6 centimeters from the nail tip, is described as U-Schenkel-Sprois 32/6cm./8mm, The double^nail with an inclined plane is a v~ry stabile syst°m, that reason it must always be in all those fractures located distally to the narrowest w rt of the marrow cavity, the fracture li of which re more eblioue or, in those which have a tendency to angulation because of a breaking off of bone particles especially at the front edge. The turn-spread nail ilso consist j cf v;o nails. The outer nail corresponds to a norm .1 nail, The ' inn nail has a round ernes s ction and is in its longitudinal profile S-shaped (ill. 19}, The lower part of the nail is introduc'd in' such a way that its concave part points w aa" front. After that the nail-is to be turned by 180 degrees • :w:n.3 of a screwdriver. The outer nail is driven in wh le the aimrr nail is secured by a crcsspin or a supporting wire against an;' further penetration. Illustr. 19 The turn-spread nail is suitable only for transverse fract- ures or short oblicue fractures the fragments of which stand one upon the other as soon as they are bolted. This nail can act only as a bolt, because the inner nail lies comparatively loose in the marrow cavity and because of its great'r elasticity it does ntt offer sufficient resistance to a stronger lateral press- ure£lll. 2t). Nails f»r_the_ Arm above tjhe Elbow, They correspond to the leg nails for children. Illustr. 20 Wedge for the Arm above the .Elbow and_the,..LegJ. They are intended for fractures proximo- to tao middle ox these bones. For the present they can only be made to measure (111, 21) Illustr, 21 Nails_for_the_ForeaTm, The cross section is V-shaped. The nails are furnished straight and are arc shaped by the surgeon according to the bone form (111, 22), Illustr. 22 Lengths: 10 - 24 centimeters at intervals of 1 centimeter. Diameter (for the above noted lengths): 3, and 4 milli- meters. The special nail for the ulna fracture (near the elbow-joint), has a length of 20 centimeters, its diameter is 6 millimeters in the third near the head end; in the third near the tip, however, it is millimeters. The corresponding tapering is in the middle third. The last three millimeters of the head end are flat and broad. The g u i d e red, For the thigk: cross section round and solid, for children’s thigh nail three-edged. Lengths: 3 and millimeters. (111. 23). For the Firearm: Length: 25 centimeters dia- meter: 2 - 2-J- centimeter. 111. 23 The Awl. The bayonet-shaped awl, the tip of which is square-edged* has proved most suitable for the leg, (ill* 24)* Illustr, 24 Nail sets^ The nail_set for the thigh nail is simply a piece of a thigh nail 20 centimeters long. It is put upon the thigh nail when the nail head end has not yet reached the head end of the guide rod and the rod must hot yet be removed, / Nail set II f«r the thigh nail fulfils the requirement that the nail should project out of the great trochanter. The hollow case slipped over the nail is pressed firmly upon the bone. The bolt is then pressed firmly upon the nail he id, bolt which can be moved in the hollow is graduated at its upper part. The number of centimeters the nail stands out at the time can be read at the top. The hollow case is wide open on one side so that it and the b»lt can be firmly gripped in order to prevent the hollow case from slipping back at the moment of the stroke (ill. 25). IT ustr, 25 The nail set for the plain leg double nail facilitates the driving in of the nail because it renders it pos ible to direct the stroke according to the chief direction of the principal axis of the nail. By this a strong springing of the nail is orevcnt- ed (111. 26). Illustr. 26 A metal bolt of 10 centimeter in length with a diameter of 1,5 centimeter serves as a simple nail set for the last strokes upon the leg nail the forearm nail and the upper arm nail. The split hammer is the bcst tool for striking out even firm- ly fixed nails. The force of the stroke is transmitted to the nail directly and in a straight direction. For the big nails the long hocks are used, for the small nails the shorter ones (111, 27). Illustr, 27 Pull and stroke are combined in the tool constructed by KUFNTSCHFR. To prevent any chipping of the bore at the point of contact a separating disk acting on a ball and socket is used. The threaded spindle with a spring acts as a "relay11 which has to be reset aft-r a couple of hammer strokes. The in- set must be inserted when the nail stands out of the bone for some distance. For removing the second piece of a straight nail, broken in the marrow cavity (thigh, forearm) the same hammer is used togeth~r with a removal hook of a length of 75 centimeters, the so-called nail_catcher (ill, 28), It must be introduced into the marrow cavitv at the site of the nail insertion. After Illustr. 28 having reached the broken nail it must be pushed into its cavity carefully so that the hook finds sufficient space in the open side of the nail. when the hook comes out at the ->nd of the nail it must be turned by 120 degrees (not by 180 derrees, see 111. 29) and the nail must then be extracted or struck out. For the a- during insertion b- extracting of the no.il Illustr, 2S ulna nail the nail catcher is slightly and elastic, and is ihtroduced by the hook closed side of the nail. The hook catche the nail tip behind at the moment when it leaves the nail, A ro- tation is not necessary in this case. The extraction hook is suitable for nails which can be re- moved easily (upper arm and forearm(111,30), Illustr. 30 The nail bender (111. 31) permits the necessary shaping of the nail (with the exception of thigh nail) without da- maging the polished nail surface an event wh'-ch may happen very easily when using pincers. Illustr. 31 The comparison-scale (ill. 32) serves to ascertain the diameter of the marrow cavity. It is attached to a belt with a balancing weight and is put upon the limb before the first X-ray in all cases of fractures which to be suitable for nailing in such a way that the scale is on a level with the middle mf the bone. The central beam of X-rays has to be half- way between the bone and the scale. bach grade corresponds to 2 millimeters. Illustration 32 The isolating .wire frame for the thigh and the leg serves to facilitate the asepsis. It erects a "wall" 4 centimeters high between the aseptic and septic parts and leaves open only a seal1 area of the skin for the small operation wound. The re Jo si t i.o ning apparatus (MAATZ) is used for the re- positioning of the fracture fragments * Practically it is a lengthening of the arms of the operating surgeon. The apparat- us stands on the floor without any connection with the extension table. Since the forces acting upon the extremity nearly neu- tralize each it is not necessary to fasten it to the floor. The floor friction is sufficient. The rough adjustment brings the grasping arms to the desired level. The movements of the lateral levers are transmitted by a universal joint to a gear mechanism in such a way that strength is saved though the extent of motion is less. By of this apparatus interlocked frag- ments may be fr~ed, moved around each othr and put together in the most favorable position for the fracture as reouired for long; oblique fractures and spiral fractures. The crepitation of the bones touching each is transmitted so well by the instrument that it is imparted to the hand of the manipulator. Each plane can be fixed by an adjusting screw. This advantage is of the ut- most importance especially with transverse and over-extended ob- lique fractures. The semicircular jDadj; can be changed, according to the thick- ness of the extremity. The leather strap which has two big holes must be pulled over the ends of the pads in such a wav that the leather strap and the pads form a complete ring and that this ring is formed at any diameter half by the pad and half by the strap. In this way a firm grasp of the extremities is achieved, The straps must be tightened very firmly because any slackening will diminish the effect of the manipulations too much. The pads are mounted on a joint that may be turned to that they stand nearly vertically to the axis of the leg or the arm. Thus the fracture site remains uncovered for the X-ray, Fractures that can be reduced only by using very strong force (i.e. supracondylar femur fractures) cannot be replaced in the normal position with this apparatus. In that case it will be better to use a regular block-and-tackle that can be attached to h*oks in the ceiling and the wall. Other adjusting devices have been constructed by HERZOG, LINSENMEYER, ZOTTL and WITTMOSER, 2* THE indication__and detailfd_statements ABOUT_THE__AFTER^TRdATMENT~ The following possibilities for the use of the marrow nail are given. The indications may be: 1# Every fracture suitable for the nailing operation may be treated with the marrow nail, or 2. the procedure is reserved for those patients who are particularly endangered by confinement to bed, plaster cast or extension treatment, or 3. those fractures which present great difficulties in treatment may be nailed, such as plain transverse fractures wh:ch often heal in lateral displacem nt and which show the tendency to refracture, or 4. the nailing treatment may be used for those fractures that may be replaced in the normal position but the fragments of which can by no oth°r method be kept in the proper oosition and which slip again and again. These fractures would otherwise re- auire an open fracture treatment. Here the indications for marrow nailing are the possibility of avoiding the opening of the fracture with all its dangers and disadvantages. These cases are not un- usual . 5. A special indication is given fsr spontaneous fractures in the cases of tumor metastasis. Strange to say the majority of these fractures are subtrochanteric, • The patients concerned must be treated with plaster cast and confinement to bed for life because these fractures do not tend to heal, not even by frequent irridiation or other forms of treatment. With the marrow nail treatment these patients can at least stay in bed without pain, in some cas-s they may even become ambulatory although there is no hope for a bony healing of the fracture. In this way every surgeon who wishes to become familiar with the use of th6 new method has a rood chance of starting with the strictest indication possible and extending the indications constantly with growing experience. Since its introduction the new method has proved its value to such an extent that the dis- position to treat every suitable fracture with the marrow nail can be considered as justified. The ouestion "Which fracture can- not be cured conservatively?" considered under the influence of the marrow nailing leads to the further ouestion "Which fractures cannot be nailed?". The indication is different with children. the fractures will heal very ciuickly, stiffening of the ioints neod not be feared, danger by cast or confinement to bed is of no serious importance, minor defects in the position are compensated for on to an astonishing extent. The indications may be the more strict the younger the little patient is. With the baby there is only the smooth transverse fracture of the thigh bone which does not regain the normal position in a suspended po- sition. But ,the oldac. the child is the easier it will bn to decide for nailing a fracture in order to shorten the time of confinement to bed, the stay in the hospital and the duration of the disease. None the less the indication will always be more strict than with adults for reasons mentioned above. Technically it is ouito possible to nail even babies. The following statements are based on the proposal to nail every fracture suitable for the marrow nail. The ouestion whether a fracture is suitable for the marrow nail operation or not depends on the shape of the bone tube, and the shape of the fracture. The situation is somewhat complicated in many cases and the gradations of suitability vary so that it seems to be advisable to distinguish the following groups: "very suitable" "suitable" "less suitable" and "unsuitable". It must be emphasized that these general directions are given with respect to the present stage of the marrow nail technique. It may be hoped and is to be expected that with the further develop- ment of this method' the general conditions will improve greatly, At present it is advisable to apply a strict standard because it concerns a treatment which is meant to replace other well-tried methods, Totally excluded are all fractures involving joints because the nail may separate fragments which still are in fair position, while the nail may not find sufficient hold and the joints are endangered by the introduction of the nail. An exception is the T-sha ed fracture in the femur section of the knee-joint. It will oe dealt with separately later on. The suitability of a fracture for the marrow nail operation depends not only on the extent to which a nail may be fixed se- curely in the bone ends, considerations may be of great importance as for example, with supracondylar fracture. Here the conditions for an ideal nailing operati. in are extremely poor. The nail $uts out of the wide marrow cavity of th'*' distal fragment as a bolt because its diameter is much too small. Additional support byaplaster cast is required. But, nevertheless there is no better method of treatment, because the fragments cannot be fixed .together so well by any other method. Therefore it is necessary to discuss the indication in every particular case with regard to the limbs and types of fractures. The circumstances, on which the suitability of a fracture for the marrow nail operation depends determine also the method of the after treatment. Therefore we must discuss the essential facts together with the indications in order to prevent any unnecessary repetition. Nevertheless, the general directions will be dealt with in a special chapter of this book. A, THE THIGH. The femur of the adult resembles most a bone tube which has the same diameter throughout, A straight nail can be introduced from the trochanter. That means that the conditions are very favorable r aJL Pertrochanteric fracture, ,T unsuitable” because the head end of the nail does not find sufficient hold in the upper part of the trochanter. Suitable for treatment with the Y-shaped nail (111. 16). O hi- Subtrochanteric fracture "very suitable” Secure against bending of the axis. Slight lateral displacement can be prevent- ed by a conic nail. Secure arainst rotation because the nail finds a firm hold in the spongiosa in the proximal sector because of its V-shape, while in the distal frag- ment conditions are favorable for sufficient friction. Additional support not required. Height bearing not advis- able until the lapse of three weeks. With a premature weight bearing the nail might work itself out laterally through the spongiosa of the proximal fragment. w Y.SI s e__and_o bl i£ue_ f r ac t ur ejs # The short fragment of which has a marrow cavity suitable -for nailing and which is at least 8 centimeters long, "very suitable” Secure against angulation lateral displacement and shortening. The danger of rotation depends on the amount of friction between the nail and the bone and is less with a jagged fracture or with an oblioue fracture and which is greater with a smooth transverse Illustr. 33 lirastr.34 fracture. Additional support is retired only if rotation occurs. Weight bearing is possible immediately, (ill, 35) d) Spiral fracture "suitable" Secure against bending of the axis and lateral displacement. Shortening occurs simultaneously with rotation faults because the smooth fracture ends will slide past each other with weight bear- ing, (111. 36) Additional supp- ort is required only when rotation occurs. Weight bearing not before the lapse of three weeks, e) Comminuted fractures 1, The largo fragments_are intact in their full leng- th in one, arean. "very suitable” Lateral displacement.js impossible. Se- curity against angulation depends solely on the stability of the nail. Shortening is impossible because the large fragments are intact in their full length in one area. Rotation ouite possible because it is pre- vented only by the friction between the nail and the bone, as the fracture is not jagged. Additional support necessary only in case •f danger of rotation. Weight bearing possible immediately, (ill, 37) 2. Intsft bone tube in thie third fragmentf "very suitable", conditions as under e,(111,38) 3, Comminuted compound fract- ure , "suitable" Secure against lateral displacement. Protected a- gainst angulation solely by the stability of the nail. Danger of shortening and rotation is very great because it is prevented only by the friction between bone and nail, (See also: gunshot fractures) Illustr, 35 Illustr. 36 Illust? « 37 Illustr.38 Additional support by wire extension probably- reciuired. Weight bearing after a sufficient for- mation of callus, (111, 39) .§uoracondylar_f r act ur e "very suitable" because it is difficult to treat by other methods. The nail prevents the small fragment from slipping out of position. Small displacements are possible in eve- ry gradation, because the nail lies as a bolt in the marrow cavity which is too wide 4 A reduction as ideal as possible should be striven for. Additional support (high plasto* cast for the leg, sometimes only Braun*s solint) always required f#r four weed Weight bearing possible after six weeks* £l_Si2E tii^condyl^r^opaxatioHjL (Y-fracture) "Unsuitable" for the simple marrow nail. As it is very difficult to treat such a fracture by the methods hitherto custor ary and as success is uncertain, the treatment with a special nail is recommended, The younger the individual conc'rr. ed the less favorable will be the con- ditions for the nailing of the th: gu, since the marrow cavity shows a fair- ly marked narrowing in the middle of the shaft during the years of growth. This fact has to be taken into con- sideration as regards the indications and the after treatment. The fractures suitable for the marrow nail operation generally are limited to the middle third of the bone. This fa.ct is not of so much importance in children because most fractures are usually situated in' that third. Illustr. 40 Illustr* .39 Illustr. 41 B. THE LEG In the leg the situation is comparatively much more un- favorable than with the thigh, even with adults. The marrow ca- vity shov/s a co-'1 siderable waist-like tapering in the middle. The nail must be somewhat bent because it must be introduced later- ally, At the same time the nail has to be flexible and elastic to such a degree that it may prove a great disadvantage for the stabilizing of the fracture. As the nail must be introduced later- ally in front, its curve must be concave shaped to the front. Leg fractures lying in the middle third of the shaft nuite often will show a tendency to recurvation. The nail which must have > sim-liar slight curve will often not be strong «nough to counter- act the forces of recurvation sufficiently. Therefore this most The fragments are protected only against rotation and not against the other possible kinds of displacement. During the insertion the nail tends to push the lower fragment backwards. It is very difficult to prevent this dislocation and in many cas«s it will be impossible to counteract it even with an ex- ceptionally wide curved radius in the upper section of the nail. Even a successful repositioning; of the nail will not protect the bone cylinders from sliding off again, Even if the frag- ments do not slide off a stabilization by the flexible nail will always be t *o slight because the latter lies almost “loose" in the marrow cavity so that it cannot neutralize the disadvan- tageous forces existing in that part of the bone. The forces disadvantageous to bony healing, are very great. This is proved by the fact that compound fractures near that place are often observed. This fracture is suitable for a nail- ing operation by the use of a wedge which must be driven in bet- ween the inner and the outer nail. By this a very good "bolting" is achieved so that the fragments stand firmly together, b) Tfap ence between tfyc dipmeter of the marrow cavity and the thickness of tile nail is smaller than the thickness of the bone mantle^ al. .Transverse fractures /in the upper third). (m - k) "less suitable" A sliding off of the fragments is impossible (111. 44). It is protected against rotation in the upper fragment by a lateral nail bed, in the low- er fragment by the friction between the bone and the nail, and in addition by any bone"teeth"® The reason for its being "less suitable" is the fact that nail, owing to the insufficient friction between it and the bon'', do~s not stabi- lize the fracture sufficiently at the fracture site to guarantee a quick bony healing without additional fixation. Plaster cast is recuired. Weight bearing with the plaster cast is immediately possible, A safe nailing may be achieved by utilizing a wedge, driven in between the inner and the outer nail, iJL Oblique fractures (in the upper third) (m-n marrow cavities. In that ca it does not make any difference, the "racture lies. The marrow cavity of the ulna as well as that of t radius arc shaped conically. Only v/ith a long nail wil1 .'it be possible to achieve a sufficient friction with the bone. Even if the forces leading to a displacement are very slight it was surprisingly often observed that' short nails inserted into the olecranon and at the styloid, process will work out. A bending of the axis is almost impossible, since nailed fractures support one another. Some examples: §,) fpaqfrurp l.Qt of tjhe..foge^riTi "very suitable11 It is well protected against al1 kinds of displacement by the friction between the nail - and the bone, Hirth-teeth, by the standing one upon the other of the bone tu- bes and by the fact that the two bones support each other, Additional bandage not Exercises immediately oossible, hi- of tfoo foreprm near the elbow-joint "very suitable" Nailing of the radius in a typical way. For the ulna a specie1 nail is required which must be considerably strengthened in the third near the head end. The reason for this is that strong bend- ing forces will be active with this fracture which must be neu- tralized exclusively by the nail, A short thick nail is not so 111 62 Illustr. 63 Illustr. 64 suitable because it shows the tendency to wander out of the bone because of the slight frict- ion in the long fracture piece. Besides this the nail end must be very broad (see page 19 ). The fracture is well protected against all kinds of displacement. Additional bandage not required. Exercise possible immediately, E. COLLARBONE The indication for the nailing of the clavicle will be given only occasionally. In nearly all cases a purely functional treat ment will lead to a nuick and satis- factory healing. If the surgeon does want to nail in of that It must be taken into consideration that vn r- few collarbones have a marrow cavity big enough for the nail and that in cases it will be necessary to . drill hole into the. bone for the nail. o.n very few cases will the r -position be possible without an exposure of the ir .cture site, 4A.§UITABLE-TIME-AND_PRELIMINARY_TRFArMENT. As a rule the nailing operation will be the easier the ear- lier it takes place the accident. In any case a shortening of the muscles must be prevented. If this does occur it might be impossible to obtain sufficient extension, in spite of the applic- ation of great force, to permit the reduction of the bone ends, especially in a fracture of the femur. The reduction will be most successful within first few hours aftrr the accident. There will then be no difficulties at all. If the operation must be postponed for a couple of days, a wire extension is to be re- commended for the thigh fracture. The traction apparatus is in- sufficient and freciuently causes ulcers by the pressure. This will happen very easily, because the most freeuent cause for a postponement of the nailing operation will be the shock effect upon the patient, The patient must not be nailed during shock. This corresponds to the general rules of surgery, of course, Even shortly after shock the patient must not be nailed, because the indication for the nailing is never urgent. As experience has shown the danger is very great even for some time after re- covery from shock. The entering of small particles of fat into the lungs must be expected (MAATZ). It is, however, so negligible that with patients who are in good conditions of health, that It never appears clinically after the nailing unless there are additional severe injuries. With endangered patients with multiple fractwrr 1 however, it may prove fatal. Therefore, an additional operative shock and an additional fat embolism must be prevented. If an un- usual bruising of fat-tissue of th* thigh (kind of trauma, marked edema of all the soft parts with large contusions) is to be Illustr. 65 Illustr. 66 Marrow-Nail in the c oll.ar bone supposed the patient must not be nialed even if he does not show- signs of a fat embolism. During the nailing operation the tissu must be moved more or less violently and handled. By this a massive displacement of fkt may oocur* The insertion of the for- eign body into the marrow cavity wil1 probably be only of subor- dinate significance. The closed thigh fracture will therefore bo treated terapora rily with a wire (skeletal) extension if the nailing cannot be performed on the day of the accident* f he adjustment of the fracture ends thus b^?ng made easy the surgeon will save much time and trouble by this comparatively small operation, while th° oeriod of waiting is tolerated much better b~r the patient. If an old thigh fracture with considerable shortening must be operated, it will be indispensable to apply first a wire extension. Even if the fractures are many months old, and if the fracture ends stand far from other an adjustment of their lengths v/ill be possible in a few days by the use of the skeletal traction apparatus. The leg, the forearm and the arm above the elbow may be con- served by applying the Kramer or Volkmann splint. An early nailing is recommended also for the leg. It v/ill be easier to nail before a more marked swelling sets in. Blisters caused by the nailing will occur only seldom and the earlier the fracture is ideally reduced by nailing the less freouently will i necrosis by pressure occur. 5^._JARC0SIS The kind of narcosis to be used depends on the expected duration of the operation and the necessity of obtaining complete relaxation of the muscles for the reduction of the fracture. In the thigh a neiling will be impossible without sufficient re- laxation of the muscles. For this purpose th~ spinal anaesthesia will be very useful. Children, juveniles and adults under 40 years are to be treated v/ith ether narcosis only when the spinal anesthesia is considered to be an unnecessary daugr to pati'nts of that ago. In the leg, the forearm and arm the elbow Evipanrausch (short 1 vipan anaesthesia) will usually suffice and it may be extended to an eth-r narcosis if necessary. Children must be treated only v/ith As a return of thp tonus of the muscles is to be desired (see after-treatment) the narcosis shoul not last any longer nor should it be deeper than necessary. All these considerations have to be taken into consideration, when the question has to be decided to what kind of narcosis should bo applied. It will, of course, be influenced largely by the cu- stom of the hospital concerned. So far only nailing of the forearm has been attempted under local anaesthesia. Infiltration at the place of the insertion of the nail and fracture site suffices. Our knowledge of the s sitiveness to pain of the marrow cavity is still rather insuffi- cient, Some patients have no trouble at all, oth°r complain about intense pain. # The proper arrangement of the patient should make possible a good extension, permit easy insertion of th<= n iil and leave room enough for the X-ray and the repositioning apparatus. Thus a free suspension of extremity concerned will be superior to the position in extension splints. Hooks in the walls and ceiling and block-and-tackle are very useful. Too much attention cannot be to the proper arrangement of the patient in order to obrain a good extension. Before speaking of the different positions and the various devices and gadgets reouired for the several bones t'v fundamental technique of the reduction of th~ fracture must be d s cussed. A reduction by hand before the fluoroscopic screen may je possible in some cases. but it cannot be a prolonged procedure because injurious effects upon thrt hands would be inevitable. Apart from this there are which require a lateral force foi the reduction of such strength as cannot be achieved by hand. Many difficulties will be encountered1 in maintaining the right position in the one plane during the radioscopy and reduction in the second plane if all this is done by hand, Therefore special tools are required. The simplest device of course are two straps which may be held in the appropriate position by or more persons (traction and counter traction). In many cases thn sur- geon will get along with this arrangement very In order to be successful in every cases better tools wil1 bn reouired. The above mentioned repositioning apparatus has proved to bo groat help. The handling of it is very simle. 1ts grasping hands are practically lengthened arm of the individual car ying out the reduction. hen reduction is obtained in onn plane the position may be held by a regulating screw while the fixation in the second olane is accomplished. The apparatus does not ~rork with rough force but allows a skilful loosening of jammed fragments and free manipulation o-f the fragments in order to achieve the most favorable position for the reduction of the fracture. T-qg manipulation will be decribod in detail under the heading "Tools'-. This apparatus will be unsatisfactory only when sufficient ex- tension of the fracture cannot be achieved, and in cases where great force is reouired. In these cases or, if a repositioning apparatus is not available, a block-and tackle will be a veyy great help, A femur fracture near the knee-joint may reouire a strong lateral force to align short fragments. Often this force may be achieved only by using a block-and-tackle. If a sufficient extension of the fracture cannot be achieved the reduction of the freture ends can sometimes be attained only by a strong bending at the fract- ure site on the principle described under osteotomy (ill. 67) The reduction of the fragments should not be undertaken until one is ready to in- sert the guide or the nail into the distal fragment. Thus an unnecessary ~nrolonged mani- pulation of the soft parts is avoided, apart from the fact that in some cases the inserted nail may well serve as an aid for the reduction, A fracture is to be operated with the patient lying on his side (111. 68), The lit. fcl 40 patient must rest on the sound leg. The hips are bent a little. This posture makes the extreme adduction of the thigh unnecessary. This arrange- ment makes the nailing possible ev- en in abduc- tion. This will be of special sig- nificance in the case of a stiffened hip- joint. The legs are spread. In t his way it will be poss- ible to apply various forces, ts reduce and to X-ray comfortably. On almost every extension table one can improvise such an arrangement, The essential factor is that both legs are attached to the same extension bar, or , that both extension bars may be moved to the same side of the table. It is recommended to construct a particularly broad pelvis supoort which allows a lateral posture of the pelvis. A stockir bandage stuffed with cotton-wool must be wrapped round the hip " the injured thigh for counter-traction. the interposition c spring scales it must be fastened sideways on the upper end of t'• table. In this way one can measure approximately the force of the traction produced at the fracture site. Mth a correct pre- liminary treatment (early operation or previous putting of the fracture in a wire extension) 60 kilograms will seldom ha.vo to be exceeded. The traction itself is attached to the foot by a canvas anklet. One of the X-ray tubes must be put between the legs and the other at the stretched side of the thigh so that the sur- ge.on may work near the bent side. The reduction is then done according to the technicue descri. at the beginning of this section at the very moment when th° guid. rod is to be inserted into the marrow cavity of the oedistal fragment. In special cases another tochnioue must be applied. With fractures near the hip joint only one arm of the reposition- ing apparatus is used. It guides the long fracture oi~co whilst the short one is guided by -.he surgeon with the inserted nail. With comminuted compound fractures the use of the reposition apoa ratus and block-and-tackle at the same time may be desirable in order to provide sufficient control for the central fragment. In order to keep accessible on all sides it should also rest on the extension table. The patient lies on his back. The injured leg is put into a leg support. The knee-ioint is bent by about 45°, the leg is extended horizontally. The leg support lies near the thigh with its longer section in order to kr ep the leg free. Counter traction is not recurred. The traction apparatus is attached to the leg with anvlet made of canvas. The X-ray apparatus as well as the reoositioning apparatus may be readily used. The sound leg hangs down vertically. Illustr, 6B If a repositioning apparatus is not available or if the reduction requires stronger force (old fracture, application of a spread nail), a modified Boehler-extension apparatus may be used with excellent result. The apparatus is supplemented by some longitudinal bars to which movable hooks are attached. Small anklets and block-and tackle allow an easy and secure ad- justment. They can be worked in four different planes (gene- rally only two will be required). To overcome a strong resist- ance which may be encountered with old fractures th:s method is to be preferred to the use of the repositioning apparatus. The use o£ jthe gluoroscope screen is somewhat difficult. A small screen is reouired. The requirement is based upon the special situation in the leg nailing, as d°alt with in detail in the chapter "After-treatment", The_arm_above_the_elbow also must be nailed with the pahi-.n in a sup$ne position. The patient lies on a firmly fixed strel cher, the shoulder-joint of the injured arm touches the end of the stretcher. The traction ap 'aratus is attached to a wristba... The arm is extended in the wlbow joint and stretched away from the body at an angle of 45%. The counter traction apparatus is attached to the armpit on the same side by a broad padded linen strip. The belt runs across the breast and back in a direction opposite to the pulling force. Gen-rally the reduction will be made easy by using this apparatus. If the place of insertion of the nail and the fractn site> are close togeth r it is recommended that the arm be drape:1 aseptically, A smal1 hole in the cloth keeps the skin free at the operation site. The operating surgeon and his assistant work aseptically. Sometimes the reduction is done by hand. In that case the reduction may be facilitated effectively by some side traction apparatus, attached and tested previously. This procedure is described, in detail for th«= forearm, because it is of greater importance for it. The asepsis of the assistant will be of great advantage in the nailing of the arm abovo the rtlbow, because the help cf an assistant is needed especially for drilling a hole In the bone. The reduction of the_forearm ipust be done with special care, since, owing to the lack of sufficient space a maximum uti- lization of the space available is required. As good facilities as possible should be available for the application of the re- positioning apparatus, because the adjustment of the fracture en is very difficult in many cas^s. The patient lies on his back on a firmly fixed stretcher. The arm above the elbow is extended at right angles to the body. The elbow joint is bent at right angle. The traction apparatus grips all the fingers by en-called "devil-finger". The pulling force is distributed equally upon the fingers by threading a cord to and from through small wheels at the rings attached to the devil fingers and the traction apparatus. The block-and- tackle then attaches to the handle of the traction apparatus. The counter traction apparatus is attached to the arm the elbow by a broad padded linen strip..and traction is exerted in the direction of the body axis towards the head. An additional counter traction is applied to the foot of the op "osite side and anklet and traction is exerted in the direction of the body-axis towards the feet with a block-and-tackle in order to fix the body of the patient who will stretch when traction is applied at his hand. 42 The block-and-tackle applied to the hand works parallel to the body axis in the direction of the feet. If the sO-called "Goose-neck"(X-ray apparatus *f Pohl) is not available, the tube of which can be inserted into a narrow space the arm must be moved around maintaining the 96° angle at the elbow joint until there is sufficient space for a larger X-ray apparatus so as to make a horizontal radioscopy possible. He who forgets the counter traction at the foot will gradually drag the whole body of the patient across the taut counter traction apparatus at the olecranon and extension will be achieved only for a few seconds. The repositioning apparatus will be used with excellent re- sult in the case of a typical forearm nailing i.e., when the nail must be introduced percutaneously at the olecranon and at the styloid process radii and if the fracture lies in the central tb:’ If the fracture lies near the elbow-joint or near the wrist it is recommended to use auxiliary straps which may be worked by hand. They are applied to the skin with the varnish ,fMastisoln before the beginning of the operation. The forearm must be stretched well and securely in order to prevent a lateral dis- placement. Their suitable direction must be tested thoroughly. They take ef ’ect at the v°ry moment the guiding rod is inserted into the marrow cavity of the second fragment. The time expend- ed on the preparations will be saved during the operation. Ex- posure of the fracture site commonly to be observed especially with the forearm may thus be prevented. An interposition of soft parts must no$ be a justification. 7^FLU0R0SC0PY It may seem surprising that a whole section of this book should be devoted to fluoroscopy. Observations made both in our own and in other hospitals, however, led to the con- clusion that it is not superfluous to do so. 'hen the technieue of the marrow nail operation is adopted by a hospital so many demands will be made on the attention and the energies of the staff by technical Questions concerning the operation that gene- rally the fluoroscopy will be somewhat neglected. In compar- ing the different precautionary measures observed in the X-ray stations with the methods adopted during the marrow nailing the latter will seem positively ruthless. This almost grotesoue difference can be explained only by the fact that people no longer realize sufficiently the injurious effect of radioscopy upon the body. To the patient the danger is certainly not so great because the operation is a solitary event in his life. The operating surgeon, the assistant and the rest of the staff concerned, however, are exposed to a great risk nearly every day, just as with cases reri radioscopy. But the patient too may come to grief. In some difficult cases the reduction may require as long as two hours, especially if it is done by unskilled personnel. In many cases the patient will be exposed to the X-ray beams far too long. The tube is generally brought rather close to the limb in order to get a distinct picture on the screen. Thus the patient may often get an overdose so that the skin may be hurt. Therefore the operating surgeon is urgently reeuested not to overlook this fact because of the great technical difficulties involved in some cases. 43 Besides this we must not forget that are some X-ray apparatuses with which the covering of the tube heats up after protracted use, and this may lead to thermic burns of the skin. The sound thigh especial’y is greatly endangered because it lies close to the tube during the fluoroscopy. An insulating layer, f. i. made of cellulose may prevent this* The operating surgeon and his assistant are certainly en- dangered much more* The war conditions will not allow for some time yet a joining of all transportable X-ray apparatus with the fluoroscopyc scr or their bring furnished with the reouired protective screen (made of lead sulphide). Many beams will reach persons near the observing doctor. Therefore it is ouite wrong to use a screen as small and as light as possible for reasons of convenience. The larger screen will afford a better protection. The situation is favorable, when the tube can be brought so close to the fracture site, or tohen the screen is so large that no beams can pass its rims. Nevertheless, the number of secondary beams will remain alarmingly great* 'the following directions are to be given under the present circumstances: The time required for the fluoroscopy must be as short as possible. It would be advisable to instruct the technical assist- ant to flash up every view for one second only. For a skilled individual this time will suffice for his orientation. For this purpose, good adaptation is required. It is ad- visable to operate in dim white light or better even in red light. AH manipulations of the operating surgeon are possible in dim light. Tube and screen must be brought so close together, or the light should be turned off so much that the beams reach only the screen. If necessary a small diaphragm ma"e of lead should be constructed. The creen must be carried with the hand well protected by lead gloves. All persons working near th~ patient should wear a lead apron. A transportable X-ray apparatus is absolutely required. It is still better to have two, because one may not wor£. If the surgeon does not want to expose the fracture it will often be necessary to interrupt the operation. The easiest way is to nail the leg without X-ray control. This should, however, be done only if absolutely necessary, nor will, it always be successfull, Fspecially suitable are the so-called uSehwanenhaelseM ( swan’s necks) furnished by the firm of Pohl. They are a small, efficient X-ray apparatus. Thanks to the small size of the tube and to three joints attached to the apparatus, the tube may be inserted into very small spaces, no loss of space being caused by the support of the apparatus. M 8a_ASEPSIS The asepsis may be endangered by the fact that one has to work in the dark during the fluoroscopy and above all by the smal~ space at one’s disposal. If the fracture is reduced by hand the assistant may work aseptically too. In that cas° the whole secticn of the limb must be wrapped up aseptically. The asepsis is con- siderably endangered by the fluoroscope or the fluoroscopic screen , Therefore a elean separation of the space required by the operating surgeon and the reducing surgeon is desirable. This separation is absolutely necessary if reduction is not done by hand. The more exact it0is the easier it will be for each to keep within his limits, therefore it is recommended to erect a separating "wall" (in the proper sense of the word) , The wire frame which is covered with a cloth has a separating wall 4 centimeters high near the aperture in the cloth. From outside the screen may be put against this little wall even in the dark. The cloth with the aperture and the wire frame must be pasted broadly over the trochanter with mastisol. The side on which the operatin' surgeon works must be screened off by a large cloth. During the fluoroscopy reouired for ascertaining tine position of the guiding rod a small cloth which must be renewed each time must be put over the site of the operation, since the screen encroaches on the space reouired by the •perating surgeon. control will be reouired less often as the surgeon becomes more experienced. He will then be able to "feel" his way into the marrow cavity, so that control of the position of the guiding rod within reach of the repositioning surgeon i.e. near the fracture site in the non- aseptic area, is quite sufficient. Also in the case of thigh fractures near the body a separation of the operating surgeon working aseptically and the surgeon who does the reducing and who works non-aseotic illy is to be r ec onmouc - ed, The latter has to direct only the fracture oioce distant from the body, whereas the surgeon doing the nailing directs the one near the body. The above described separation is easily achieved when the leg is concerned. The cloth-covered wire frame covers up the bent knee-joint. In the arm above the elbow and the forearm one will prefer a complete protective cover the closer the place of the insertion of the nail to the fracture site. The small space available makes a separation difficult. Nevertheless, it must be accomp- lished when using the repositioning apparatus. See chapter " Proper position of the fracture ends". The traction straps des- cribed there must be fastened before the protective cover is attached and they must be tested as to their capacity. For the operation wound, slits in the cloth 1-2 centimeters long, will suffice. Here also the cloths must be pasted upon the skin near the aperture. The auxiliary traction straps must be covered up too. Their ends will of course project from the aseptic area. The length, thickness and shape of the nail depend on the thickness and shape of the bone as well as one the position and the kind of fracture. It must he 45 success is not certain. Therefore, the diameter of the mar ow cavity should be ascertained accurately at its narrowest part. At a focal distance of 75 centimeters the average record on the plate will amount to about 1 millimeter. It is also possible to hold the probably suitable nail beside the limb, when taking another X-ray so that it will be possible to take a comparative measurement, as the distorsion will be the same (EHRLICH). To use a "Metalltrepee" (a scale made of metal) will be even better, because it is simpler. It is furnished with a strip of leather and a counterweight and is put on the fracture at the level of the bone, when the first X-ray is taken. The same record will be obtained when the central beam is put between the scale and the middle of the bone. Each grade of the scale is 2 millimeters long* The grades run in the direction of the axis of the extre- mity, In this way the bending out of line of the scale will not become a source of mistakes. So, even with the first X-ray the width of the marrow cavity can bo read. For the thigh it is preferable to us~ a thinner nail. The marrow cavity which is largely of the same v/idth affords a sufficient friction. With a nail which is too thick, however, the friction may become so strong that the nail jams. During every operation all the three different widths of the selected nail should be available (see insertion of the nail). For the_leg it is recommended to us- a nail as thick as possible. the marrow cavity is narrow only for a short distance. The nail may be driven in >ith some force because it will enter easily thanks to the elasticity of its cross soction. So there is the advantage that the nail jams in the marrow cavity to as large an extent as possible. For E-.lhoyr_the_elbow it will seldom be possible to solect a nail corresponding to the diameter of the marrow cavi- ty since its latoral insertion is rather difficult in this case. If, however, the surgeon selects a nail which is some millimeters thinner than the measured width the nailing will be successful and the two nails will provide a sufficient hold. If the fract- ure lies near the elbow joint the nail should be so thin that is certainly cannot jam in the tapering marrow cavity at the el- bow joint. For the stable osteosynthesis the jamming will not suffice anyhow, a plaster cast will reeuired and. the jamming to be feared would lead to a blocking caused by the nail. In such a case we have to sp*ak of a ‘'bolting”. For the__forearm__ in most a rather thin nail must be selected for a typical nailing operation, since ulna and radius have rather thin marrow cavities in their tapering parts. This disadvantage will be offset by the fact that long nails may jam at many points of the marrow cavity along the whole length of the bone. Short nail, however, inserted from the side and nearer'to the fracture site should and may be thicker. Their being inserted from the side will, however, make it impossible to utilize fully the thickness corresponding to the diameter of the marrow cavity. The hold provided by such nails is not sufficient. Here, also we have to sp-ak of a ''bolting”. Tjie_lengtji of the nail is of some importance only in con- nection with fractures near the joint, which are ,t t'l-' same time distant from the place of insertion. In such a case one must not miscalculate. vven one millimeter may be valuable or cause trouble. •46 In case of all other fractures one should bear in mind that too deep an insertioh of the nail into the peripheral fract- ure piece can only be superfluous and will never cause trouble or be disadvantageous. The length may be meaHured best on the sound extremity, at tho thigh from the end of the trochanter to the slit of the knee-joint. From the slit of the knee-joint upwards one has to subtract 3 centimeters. Au the trochanter one must add 3 centimeters for the part of the nail sticking out so that the measured length corresponds to the allowed length of the nail* On the leg one measures from the dimple between the sinews of the extensor digitorum lon?us and the tibialis anterior muscle at the heigh of the ligamentum cruciatum curis to the middle of the tuberositas tibiae. Because the nail should be at least always 1 centimeter distant from the joint plane of the foot, the measured distance corresponds to the desired length of the nail. For the bending 1 centimeter must be allowed. The leg nail should be as long as possible. Special attention must be given to the selection of the leg spread nail. chiefly matters is that the inclined plane cf t he outer nail, which pushes the inner nail backwards should have the right • * position. If the inclined plane comes to lie in that part of the marrow ty v/hich is too wide it will be in- effectual because the counter pressure of the bone will be lacking. The end of the inclined plane lies at the spot at wh ch the marrow cavi- ty is just sufficiently wide for the two nails \e.)m It will be advisable to measure it distally, i.e, fro the foot' joint plane. Since tho nail point should be 1 centimeter distant from this spot, that centimeter must he subtracted (f-1 centimeter - distance of the end of the inclined plane from the nail tip) (111.69) In the__arm__above__the_elbow the distance from the intended insertion place to the usual insertion place near tho elbow joint should be measured if the nail is introduced from above, as the marrow cavity ends there. If the head of the arm bone is to be taken hold of too, i.e., if the nail must be inserted in an upside-down direction, tho measure should be applied at the corresponding spot. For the__forearm it is preferable to take a nail as long a® possible if it is a case of a typical nailing. In thiB way the fractures are best secured. A wandering out of the nail, ob- served especially in the forearm near the elbow joint as well as at the wrist, is made impossible. For the radius the nail must reach from the styloid process as far as 2 centimeters from the head of the radius, and in the ulna from the olecranon as far as 2 centimeters from tho styloid process. Tfoe longitudinal profile of the nail is generally dctor- mined”and prescribed. Nevertheless, it is recommended to modi- fy that shape in special cases. In Northern Germany the straight thigh nail is generally used. It is easy to insert in the slightly bowed bone. In other parts of Germany the bones are sometimes bent so much that the nail too has to bo bent before its insertion. But one cannot depend on its following the shape of the bone. If the fracture site is near the greater curvature of the bone there is the danger of a splitting of the bone. Fven though the splitting of a straight thigh bone by a jamming marrow nail is hardly possible, a large piece of the cortex may ILL.. 47 easily be broken off of the greater curvature of a bowed ''one, if the nail is driven only against one side of the bone. In the leg the middle part of the nail should always be straight, in order to prevent a recurvation. It is recommended to give to the inner nail a stronger curve than to the outer nail, in order to achieve a better jamming. It should be emphasized that the surgeon who is not yet fully skilled in this operation will do well to insert the double nails one at a time. Later, when greater skill has been acnuired they can be driven in at the same time. The nail for the arm above the elbow too should be as straight as possible in its middle part. If a short fragment is entered for only a short distance by the tip of the nail that tip should not be curved. With the forearm the nail f*r the radius only should be bent somewhat at the head end. Thus no forces can act upon the wrist when the nail is once fix- ed. In order to facilitate the insertion the nail should be bent only after about two thirds of it have entered the bone. Otherwise the nail remains straight. For the ulna a nail of a straight shape is always used. (111. 71) For a fracture near the elbow-j»int a special nail is used, which is thicker than forearm nails and tapers at its end . 1 his shape is reouired since the nail should be fixed very firmly ai the fracture site. At the same time it must be driven into the ulna so deep that it cannot work itself out of the bone. The nail end must be particularly broad so as to find sufficient hold in the spongy bone, A thin nail will cut the soft spongiosa (see chapter I). III .^0 III/) I OF THE GUIDE ROD In order to facilitate the introduction of the nail into the proximal marrow cavity a guide rod is used for the nailing of the femur and the ulna. At the same time this has the ad- vantage that the guide rod can easily be pushed forward at the moment of a successful reduction in order to maintain the re- quired position of the fragments. When the nail is driven in by blows with a hammer an unintentional displacement of the fragments may easily occur. Whereas a guide rod wrongly placed may easily be taken out again* the striking out of a nail may offer considerable difficulty especially in the thigh. In a nulling of the ulna it will be a great advantage if the guide rod is inse ted into the distal marrow cavity a short way at the moment of a successful reduction. This technieue would be help- ful also in the radius. Here, however, it wil1 be advisable not to use the guide rod, in order to avo d all the stress not absolutely reeu"red which might cause a breaking of the bone at the bone at the styloid process near the wrist. 48 Both guide rod and nail must have the necessary free play, 0 nly. sui t a1'le sized instruments may be used. If the guide rod is^fixed too tight in thQ nail a jamming may occur the moment the na.il is inserted, since the na.il shows a tendency to adapt itself to the shape of the bone and must be free to bend more or less. For the insertion of the guide rod at the thigh the direction of the marrow cagity must be indicated on the lateral and posterior sides of the thigh with a skin pencil. Those lines are determined during the horizontal and vertical radioscopy. To protect the ha.nds from exposure to the X-ray beams during the radioscopy it is recommended to use a long guide rod which must be brought into the direction of the marrow cavity. The lines will then be marked upon the skin accordingly. It is obvious that the direction of the marrow cavity of only the proximal fragment is important. The lines are extended and distally for the entire length •f the thigh. If the radioscopic planes are correct both lines must meet near the trochanter and at the insertion snot of the guide rod. Nevertheless it will often occur that the point of intersection is farther in the dir-ction of the trunk. So it is of no value for determining the spot for insertion of the guide rod. It will be better to go by the palpation finding. The trochanter tip can be felt easily pvon with corpulent people, A stab incision 2 centimet rs lorn? is made slightly in the direct- ion of the head and mddially from the tip. Then the guide rod is inserted into the trochanter. One may also drill a little h#le into the bone with the awl, that the guide rod is pushed forward by hand, turning it slightly. In this way the guide r»d will not jam tio much in the trochanter, and it will be much easier for the surgeon to feel whether the guide enters into the marrow cavity, or whether it is in a wring direction or not. The direction lines of the marrow cavity indicate how the guide rod is to be held. The upper part of the thigh being covered with sterile drapes andthemarcation being covered up also, the marcation lines should be extended onto the lower part of the thigh as described above. The guide rod must not be too sharp at its tip. Otherwise it may occur that it jams laterally in the corticalis. This fact must bo taken into consideration especially with children’s bones. After a short time the surgeon will get an excellent sense of whether the guide rod is in the right position or not, so that additional radioscopy is reouired only v/hen the tip of the guide rod approaches the fracture site. If, however, the in- sertion should prove difficult, it will b-" to desist from making further attemtps and to ascertain right position using the screen, In case it is faulty it will be advisable to make another insertion. Otherwise the guide rod might slip again and again into its former position. There can e no objection to push- ing the- cruide rod forward by slight hammer strokes. As soon as the surgeon has a certain esperienen the sound of the instruments, and the success of each stroke of the hammer will indicate to him whether he is making satisfactory progress or n»t, . - - . ' • * ' ". a • * The insertion of the guide rod, may be rather difficult-, in the case of a fracture very high in the upper- pArt. of the bone. If short fragment is. in abduction i£ will.in rare. cas.es be necessary to use a strong so-called,. Schanz Dorn (a. pr>ick-.puhch constructed* by Schanz). If inserted percut an.eously. and laterally into the trochanter it. will, be a convenient handle until the , guide rod. is introduced and the reduction is achieved. If the short fragment stands in an adducted position this may lee.d to an increased adduction of the guide rod. The tip #f the guide rod 49 then shows a tendency to jam in the corticalis (outside). A vicious circle will be the consequence. The insertion of the gui d ' rod will, however, be considerably facilitated if a traction strap is attached to the short fragment before the operation, which by pulling in an outward direction will give a certain stability to the short fragment. It will not be difficult to introduce the nail into the distal fracture piece after a successful reduction. In the ulna the guide rod is inserted through an incision in the skin meter long, at that point of the olecranon wh're the extended marrc cavity of the bone is supposed to be. There the gu de rod must be drilled in or driven in. This spot can easily be detormined by touch, for, the center of the bone can be felt distinctly. Appropriate X-ray control in both planes will be a good protection against any mistake. 11. THE INSERTION OF THE NAIL By way of introduction something shall bo said about the tech- nique of nailing* Generally the surgeon wi1 1 be inclined to introduce the nail with many light careful strokes of a hammer. It is better, however, to insert the nail with some powerful strokes driving the nail well home* Thus the bone Is not shaken so much. A continuous striking upon the bon.'' will often lead to an infiltration of fat into the blood vessels (LARSRN). Besides that the striking in of nails in a workmanlike way is less tediou: and much more elegant. It is not at all difficult to insert the nail into the thigh, as soon as the selection of the na.il and the position of tha guide *nd are correct. It does not make any difference, whether the guide rod is inserted first into the two fragments or into the proximal fragment after the distal fragment has been pierced slightly by the guide rod. The latter method is prefer- able only when the fracture-piece near the body is short so that it must be guided with the nail during the reduction. The nail being elastic and flexible on all sides it is of n» importance at all to which side its open part is directed, Trr resistance is fairly strong, while it is in- serted into the trochanter. As soon as it has entered the marrow cavity in the right way it will penetrate about 2-3 centimeters with each stroke of a hammer weighing about -g- kilo- grams. A splitting of the bone is not to be , feared, more likely it will get stuck in the marrow cavity. One must, however, see to it carefully that the nail is not driven in sub- trochanterally through the compacta. This may happen when the guide rod driven in by hammer strokes first pierces the compacta and reaches the marrow cavity only after- wards, In such a case there is danger of splitting the bone (ill. Since the degree of friction between the nail and the bone depends largely on the small ridges in the marrow cavity, it may happen again and again that the thickness of the selected nail proves to be unsuitable. If the nail tends to get stuck or if it has not sufficient friction it is preferable to take it out again and to replace it another nail. The trick is to find III ,r the shape of the fracture (ill. 77) When the turn-spread nail is used the inner nail must be introduced complete- ly first, and in the begin- ning in such a position that its distal part is concave shaped to the front. As soon as the point of the nail lies in the wider section of the lower marrow cavity the nail is turned round 180° using a screw driver. This turning must be made before the point of the nail has penetrated into the spongicsa of the distal bone section. During the insertion of the outer nail the completely inserted nail is then protected against further pene- tration by a holdfast wire or cross-pin which is inserted into an ILL 1 S' mu ILL.'/1 52 an oblique hole *. - at the nail head provided for this purpose and which finds a firm hold on the tuberositas. If, however, the outer nail should show a tendency to w»rk ito'way out of the fracture-slit in front, suitable dislocation must be made during its insertion into the marrow cavity (ill. 7$). I In the arm above the elbow the nail may be introduced at tw® places i,e, near the elbow joint, cr near the shoulder joint. Both places are easy of access and the choice will depend on the site ®f the fracture. Two con- sideiiations are decisive; In the firs4: place the nail is t® find a firmer hclc: in the short fragment by penetrating the corticalis laterally. In the second place an exposure of the fract- ure hematoma should be avoided. The., fire a very short fragment should not be nailed from the side of the hema- toma ®therwise the fracture site would be exposed. Fractures in the second and fourth section of the bone should be nailed from above, all the other fractures from below (ill, 79), The upper insertion point must be exposed by an incision 5 centimeters over the deltoid muscle. The muscle must be split longitudinally with a crown-gimlet and a hole must be drilled from the side into the corticalis, so that the marrow cavity is opened up. Then the two nails can be inserted obliouely. They .should project out of the bone for l-g- centimeter. If the nail is inserted ?,roin below the drill hole is situated about 2 centimeters above the fossa ®lecrani. Here the marrow cavity may be so flat that an extension of the drill hole upwards may be required. At that spot the nail should project out of the bone for about one centimeter. If the nail inserted from above is too long it must under no circumstances be inserted too deep. Otherwise it would jam in the marrow cavity near the elbow. It thus push the second fragment al#ng before it and might cause a gap. If in such a case the nail head end should stick out of the bone too far the nail must be changed. In order not to loec t h e reduction a thin guide rc 1 must be inserted before the nail is removed ( if need be a Kirsch- ner - wire may be used), A special description of the nailing of the ulna is not need- ed. The straight nail may be inserted with the help of a thin guide r®d. At the radius the insertion ooint for the nail lies between the sinews of the extensor poll, longus and brevis. Over the middle of the styloid process radii a stab incision 5 centimeters long must be made and a tiny hole must be drilled oolieuely inuo the marrow canal with the awl. X-ray control while taQ awl is put t® the bone will ensure the finding of the right -pot. The Illustr. 78 III. 53 nail is always inserted with a slight pressure of the finger away from the wrist. The nail head end is bent before its final insertii (see shape of the nail longitudinal profile). The cutaneous wound is closed with a stitch-suture. 12^thf_aftertrfatmfnt Only a few words need be said about the aftertreatment of a successful nailing. If the oossibllity of a displacement of the fracture ends is out of the ouestien the moment of the first move- ment of the limbs and the time for th° first weight bearing depend, tn the general condition of the patient and the extent of the very painful bruising of the soft parts. Fffusions of blood into the adjacent joints which used to occur much less freouently when the old methods were ih vogue, and which are only partly caused by the forcible extension during the operation may demand a long confine- ment to bed. Whan making up a list of a number of cas-s one is surprised to find how few patients with thigh fractures can or may get up within the first week after the nailing. The number of attendant njuries among which concussion of the brain or skull fractures are particularly frecuent, is astonishingly gr~at. Though experience has shown that the aftertr»atrent does not give much trouble either to the surgeon or to th° nursing personnel the vigilance of the surgeon must be all the greater. In case of insufficient care the number of persons discharged too early only to return to the hospital after few weeks with unexpected dislocations will not be small. Malpositions must be detected at a time when they may be corrected during the "Rausch" (anaesthesia) in order to prevent a later osteotomy. The aftertreatment begins sft°r the wound has been closed i.e while the patient is still under the anaesthetic. 1. With nails less difficult to bend (leg and forearm) the proper shape of the nail must be verified using the fluoroscopic screen. The leg nails in particular may require some correction as they very often get bent in the form of a recurvation of the le during the insertion. 2. The possibilities of a displacement of the fracture ends must be examined as soon as the nail has been inserted, If a bend- ing of the axis is still possible, an additional plaster cast is urgently reouired. 1f rotation is possible, an additional plaster cast must be applied if the shape of the fracture makes is prob- able that even the return of the muscle tonus the principle of the Hirth teeth is out of the ouestion. If a fixation by the Hirth teeth may be expected a splint which will prevent a tipping of the limb until the end of the narcosis will suffice. The examination to be made the following day is decisive for the fur- ther kind of treatment. In this connection it may be pointed out that, while a large log cannot be dragged with a jerk nuickly throw' viscous mud, this will be easily done with little expenditure of force, if one only has sufficient patience. The friction between the nail and the bone shaft may be strong enough to prevent a cuick rotary movement of the unstable foot and stil1 not be strong enough to withstand the constant though small force of the twisted f o#t, 3. After any smoothening off of the sharp-edged fracture ends the effect of the Hirth-teeth must be disregarded. 54 4, After the lapse of a couple of tha nail will get ltose*in the marriw cavity., If conditions are favorable the fracture ends may slip longitudinally on the nail without -any great danger of rotation. Nevertheless the possibility of an. rotation displacement rust be taken into consideration. To avoid unneeessary repetition detaiIs about the data of the aftertreatment will be given in the chapter on nIndications" for the same factors which make a fracture more or less suitable for the marrow nailing determine the kind of aftertreatment. By way of general directions it must be mentioned that active and painless exercises will create the best conditions for the healing of the fracture and the conservation of the strength of the 3-imb. Baths and heat cradles should be freouently used. Complications during the healing of the wound must be de- tected early. It would be wrong to delay if a suppuration of the insertion spot is suspected. A prophylactical"y opened wound wil1 heal quickly, whrle a wound opened too lat« may cause in- flammation of the fracture hematoma if the lies near the nail insertion spot. A slow spread of the inflammation along the nail is not to be feared if the wound is opened early enough. The time reeuired for treatment in the hospital deoends on many external factors. Surgeons not yet fully acquainted with the new method are urged to leave the patient in the hospital as long as possible. The duration of the patient’s stay at the hospital depends on many external factors* It is important what profession the patient has and. what attitude he takes. An office clerk f.i. with a w~ll treated thigh fracture may easily go to his office at hiw own reouest a fortnight. A workman, howev i will n»t be able to go t# work again befor- the lapse of 6-8 weeks even under the most favorable conditions. Of course the clinical finding still will be decisive. It is the responsibi 11 •. of the surgeon to suggest to the patient the right attitude of mind towards his disease with regard to the new treatment of fractures and the shorter duration of the disease. The surgeon as well as the patient must get used also to the much lower rate, of sickness-benefits paid in favorable cases. Special attention must be oaid to the progress of an "asepti. inflammation of the fracture, until sufficient experience has been gained (Sudeck), Even when the nailing was successful a dys trophic phase may dwveltp (painful swelling, heat, cyanosis, perspiring skin). * Early*.weight boaring seems to bring on these complications, ' The best treatment for these are rest and the heat cradle* ll^-THF^R^IfOVAL^OF^THE^NAIL In general a nail should be removed the lapse of three months since a firm bony consolidation of the fracture must be expected by that time. An earlier removal may be advisable. Reasons for this are; 1, The fragments hav- become displaced and the nail hind- ers a new and exact reposition (anvulation or rotation). 55 2# The nail wanders and endagers neighbouring joints or the soft parts covering the nail head end. 3. The nail is bent. 4* The nail is broken. 5• ith children to attain a ou ck°r healing of the fract- ure. Suppuration near the insertion spot or fracture site, how- ever, (osteotomy, complicated fracture) are not indications for a removal of the nail. In th- case of a fistula at the nail in- sertion spot the extraction is even undesirable. It will be ad- visable to wait as long as5 possible in general since such fistular will eventually disappear, whereas an early extraction may lead to a late suppuration near the fracture site. Obviously, the suppuration will spread along the marrow cavity much more easily when the nail has been extracted. The nail must lie in the bone for more than three months if it is suspected or certain that the fracture has not yet healed properly. A late consolidation may be caused by; 1. Infection. 2. Distraction of the fragments due to jamming of the nai.1 3. Insufficient mechanical stabilization of the fracture by the nail. 4- All the other factors which may load, to a delayed for- mation of callus (Diabetes, Lues and so on). They generally known from the science of fracture treatment and need not be repeated h^re. 5. With an osteotomy performed because of a faulty posit."., after breaking the bone it may be advisable to leave the nail in for a longer term than three months since the normal balance of the muscles is reestablished only slowly. Besides this osteo- tomy wounds achieve a slower bony healing then recent fractures nailed percutaneously. The adequate bony healing of the fracture is ascertained by X-ray control. The nail may be extracted as soon as the fracture fissure is bridged over the bone sufficiently and as soon as the callus shows a satisfactory formation of the little streaks. If this cannot be ascertained beyond, doubt the nail should be left in the bone for eVon though with some fractures the fract- ure fissure is to be recognized euite distinctly and. though it mr.y be firm from the clinical point of view and suitable for weight bearing, this cannot be ascertained beyond doubt as long as the nail is in the bone. Therefore it is better to leave it in. Xai' _ inserted two years previously - not caused any particular harm to the patients concerned. though the X-ray picture shows a vigorous formation of callus at the fracture site, it will not be advisable to remove the nail while the design of the callus is soft and cloudy. Cal1us which appears soft and cloudy In the X-ray is often soft and flexible so that an angulation may occur after the extraction of the nail. Fven though a newly inserted nail may be extremely difficult to extract, nails inserted a couple of weeks or even months before, are generally easy to remove. Loose nails may be extracted under 56 l*cal anaesthesia (q.rm above the elbow and forearm) while firm- ly fixed nails should be removed tinder pan-anaesthesia. For the_th_igh the patient must be operated while lying on his side; It may be difficult to find the nail hole if the nail head is covered by a callus cap. The "Siemens" metal-finder by KUENTSCHFR is very useful for a ouick orientation* If the nail sticks firmly and if the surgeon intends to manage without a large incision of the skin it will be better to use a hammer for re* moving the nail. ./ith its help dvory nail will be easily removed. If spread nails with an inclined{plane haye been used the inner nail must alwayd be removed fiifst. If the turn spread nail has been used the outer one first, Since dotation of the inner nail may be impossible because of the marrow The nail which is deformed may raouire considerable force for its removal* A broken th?gh nail must be removed as soon as possible after this accident has occurred. The will soon go to see his doctor, because of the pain at the old fracture site. Frequently an angulation of the thigh will occur. The fracture site is not yet firm, otherwise the nail would not have broken. The break in the nail is not always situated near the fracture site of the bone. The proximal part of the nail is removed firsi, then the nail catcher (see t*ols) is introduced and as soon as it has "caught” the remaining piece the nail must be extracted or hammered out, A bending at the fracture site must first be ad- justed. The insertion of a fresh nail will not give any trouble. The finding of the nail head in the leg may be difficult, if it has worked itself deep into the bone, which will happen frequently. Here also the "Siemens" motal-finder will prove very useful. If it is not available it will be advisa le to insert a small pin before the screen or to make a small X-ray picture (a-p) with the pin in the right position for the nail head must be sought for beneath the patellar tendon. Sometimes it is perfectly amazing how many centimeters th° nail is to be- found distant from th~ place t was supoos d to be either laterally or medially. By the described procedure unnecessary injuries to the patellar tendon are avoided and the associated continuous irritation of the wound, which wi Zl occur frequently if the ion of the na: 1 in the leg is not done carefully. using a spread nail with inclined plane it must be borne in mind that the nails must be removed one after the oth°r and that the one with an inclined plane can be removed only after the other nail has been extracted. A broken aail in the leg must also be removed as quickly as possible and must be replaced by a new nail with the same diameter. In order to avoid disturbances at the fracture site the surgeon may desiflt from an exposure and extract the distal piece of the nail first -ith the nail tip. The nail tip is exposed at the ex- tensor side of the tibia, then a hole must be drilled into it and it must be extracted in an oblique plane. This is mane possible by the elasticity of the nail. In the arni_aboye the elbow there ar" no difficulties to be encountered either in finding the nail or in extracting it. In general it can be removed with a hook. An exact X-ray location of the nail head end is recommended if the nail has wandered and has completely entered the marrow cavity, and if the metaT. finder is not available. If this is the case the marrow cavity must be opened laterally and the nail must be pried out by leverage and extracted on an inclined plane. As regards the__forearm only the removal of the radius nail at the ulna which was inserted at the styloid process must be specially meni oned, since during its insertion as well as during its removal an injury to the wrist 57 must be avoided in any case. Every movement with the lever must be executed in such a way that the bone is not pushed towards the wrist and gets broken. This can be achieved with some care. It must be taken for granted, however, that the nail was correct- ly formed before its insertion during the first operation, so that it does not show any tendency to winder into the bone. The radius nail as well as the ulna nail call., be extracted easily with the hand after the first loosening strokes. Breaking of nails was observed alro with ulna fractures near the elbow joint. This fact proves how strong the forces ar° which act at that spot. Since th~re is a straight nail in the distal nail fragment may be taken out with the smal' nail catcher (see tools). 14^_MI1IIG_IN_TH^_CA£F_0F_AN_0PFN__FRACTURE The advantages of marrow nailing are thp samp both for closed fractures and for those which a primary healing of complicated wounds. Probably primary healing will occur much more frequent- ly with the new method than with th'' old on^s} sr'nce bony wounds and thos~ in the soft parts are‘immobilized better by the nail than b any other method. The new method shows the following great advan ages in the case of a suppuration in th© front.of the fracture site : 1. The fracture has been ideally fixed. A spreading of the suppuration of the soft parts to the bone is not easily possible If the bone wound is affected by the inflammation the suppurati will proceed more favorably by reason of the fixation. 2, The extremity is easy of access on all sides. So it can be watched every day and incisions reeuired can be made early. Disadvantages of the method are: 1* Suppuration may proceed along the nail. In this case it is restricted to the marrow cavity and does not affect the ben» tissue (A. W, FISCHER and REICH). Fven after the lapse of weeks such an inflammation at the insertion site of the nail may lead to the formation of an abscess under the primarily healed operatic: ’wound. After the opening of the abscess the nail acts as a drain for the often very active discharge of pus. 2, At the fracture ends crown-seauestrum or ring seouest- rum may develop. The sequestra are covered by enveloping callus which must be removed after the fixation of the fracture. For the nailing of the marrow cavity of compound fraetures the following direct- ions may be given (MAATZ and RFldH): ( The nailing operation should be done as soon as possible. A limit of hours cannot be fixed but during the course of an in- fection a nailing must not be performed. After the subsidence of the initial violent inflammation thp comes another moment favorable for the nailing. After hav- ing attained a sufficient immunization from the germs in question and after the suppuration has reached the so-called "cold stage" 58 the body appears to be able to endure a surgical intervention without any injurious consequences (see gunshot fractures). Additional injuries to the complicated wounds in the soft Parts should be avoided in the nailing. Therefore a technique similar to percutaneous nailing is to be recommended. The guide rod should not be inserted in an upward direction as :n the ease of an osteotomy. Fven if th~re :s no X-ray apparatus available the guide rod can be inserted precutaneously the surgeon has some experience. The r duction and the insertion of the nail into the distal marrovj cavity can be control1 ed by the eye. In the case of small wounds too this should bo done before the fluoroscopic screen in order to avoid opening of these wounds further, If possible complicated wounds should be treated after the rules for the primary excision of wounds with subsequent suture. If the wound is very dirty and if it is impossible to excise it with subsequent suture a debirideirteat must be performed, i, e, the worst dirt and crushed tissue must be removed and then the wound must be left wide open and covered loosely with gauze. In t the case of a formation of pockets a squcerization of the wound is to be recommended, in order to prevent a damming of the secret- ion. Apposition-sutures are to be rejected. By doing so the wound will be cleaned in the quickest way and secondary healing will be achieved. The latter can be accelerated by later nappo- sition-sutures”, Thus a progressive downward extension of the inflammation is prevented. The advantage of the accessibility of the limb must be ful- ly utilized in the course of the further treatment of the wound and one should adhere to the principle that by leaving the wound open the discharge of the secretion should be secured early. An additional cast should be avoided if possible. This, however, will result in a narrowing of the indications. The advantages of the treatment of complicated fractures with the marrow nail depends on the complete stabilization of the fragments and the accessibility of the extremity. If an additional cast seems to be required prior to the nailing there is no indication for the nailing operation. The success of the nailing of an open fracture d-pends essontailly on whether it will be possible tc achieve a stabile osteosynthesis wjth the nail. I Tn general the extremity must be fixed on a splint, as it is the rule with inflammation processes of the soft parts. In this way a changing of the bandage is easily secured. The nail should be kept in its position as long as possible, i,e. until a bony healing is achieved. Fven a very violent suppuration which seems to be caused by the presence of the nail should not be a reason for a premature removal o_f the jaai-1, at least never during the first 'eight to ten weeks. rven though V2A tr V4A steel is the least irritant metal to the human body, known so far, the body will cover it with a connective tissue membrane and will try to eject it by suppuration after an in- fection has occurred. The advantages of nailing are still so great, that an increase of the suppuration can certainly be risk- ed. The formation of callus is certainly not delayed by the nail. Nevertheless, suppurating fractures with an inserted nail will frequently not show the formation of any callus for months. In such cases it is not the nail which is responsible for it but the suppuration. It is always wrong to extract the nail as long as a displacement of the fracture may stil1 occur. This may happen even after th° lapse of some months. If a fracture 59 is well protected against lateral displacement either by cica- trization or calnus, a removal of the nail may be allowed if a particularly violent suppuration is to be stopped. The securing of the fracture against angulation must be achieved by an addit- ional cast. Our aim must, however, always remain the bony heal- ing with the inserted naili Sometimes it will be important not to patience and nover to forget that complicated fractures will sometimes supourate for months even v/ithout a nail and may not Show a tendency to form any callus at all. The most favor- able conditions are still achieved by the nail* .FRACTURES Any comparison with open fractures must be avoided. Where- as a wound excision and suture may freouently be mare with these, it is not permissible with open fractures. Owing to the peculiarity of the situation some general rules should be considered: The nailing operation should be performed as soon as possi .. One is, however, not limited to certain hours. A successful operation may be performed at any time (KUENTSCHER), Patients with violent sup uration and high fever will recover astonishingly auickly after the fixation by the nail (KUENTSCHER). Wot only a careful operation but also a v/ide opening of all the wounds is necessary. The insertion spot must be larger than usual and must not be closed by sutures but must no loosely covered with gauze. Adeauate drainage must be secured by sufficient counter- incisions, The use of a wound spreader may perhaps be necessary. In the case of gunshot wounds causing bone defects a metal spreader ehbuld be inserted into the nail slit in such a way that the bone ends are at the desired distance from each (KUENTSCHER). The spreader S'rvos at the same time to keep the wound open. The insertion of the guide rod from the fracture site is to be avoided because an unnecessary spreading of germs is to be feared an experienced surgeon will easily find the marrow cavity from the trochanter even without X-ray control. Inex- perienced surgeons should not nail gunshot fractures at all. Reduction before th- fluoroscopic screen is not advisable, since any maltreatment of thn tissue must be avoided. It will be better to widen the wound by a straight incision as explained before, in order to bn able to control the infection. ;FRACTURE Spontaneous fractures must be given special consideration in connection with the nailing operation. In most cases they are fractures of weak bones, enfeebled by tumor metastasis, in rare cases b'' osteoporosis or other factors. Curiously enough the tumor metastasis is to be found chiefly in the upper third of the thigh, i,c. at a favorable spot. In connection with th,© nailing it must be taken into consideration that in general the nail is meant to provide a sufficient stabilization for the rest of the watients life, because these fractures do not show a ten- dency (or only very little) to healing even after X-ray treat- ment. In the second place chiefly atrophic bones with a very thin compacta and correspondingly large marrow cavities are con- cerned so that customary sizes of nails can be used only seldom. In the third place the bone is comparatively soft, and in the fourth and last place the nail is exposed to little bending force, as the patients are usually elderly and frail persons. The facts given above make it advisable to a special nail„ As the nail need not be particularly proof against bend :ag the sheet metal may be somewhat thinner and the elasticity of its Illustr. 87 cross section can be all the greater. its hat-shaped cross-' section grants to it not only the required, elasticity but has the further advantage that the nail touches the bone wall only with its three narrow longitudinal plants, its three edges. During the first few weeks the bone yields to the pressure of the nail. Because of its elasticity the nail formes grooves wh: ch permit it to attain its former shape and so become loose. These grooves are deep enough to prevent a lateral dispalcement of the fragments with certainty. Fven a patient who cannot hop© for a bony healing of his fracture will be able to walk with his bones in splints without any additional protective casts. For the rest, the technique of the nailing does not differ from the usual method (111i 88 ) * In addition only the softness of the bones must be taken into consideration, especially when >. ! ■ Y vs --f d p\v f i i \ t / < //* \ \ is Vi ■ \ / t-K 1 KA ir do V ! \\\l J ihi\ i',\\v 4 f ; /f X m I p \\ i / ■ / / 1 i i* S III ,;■'// j|| i rl Pi 1 I'1 *1 | /!■:/ ! •.! 1 f ! : i Si ' | h j ; v I .« < r iilh ; ■ ui.« d* 'i !■ » A ILL .SS inserting the nail into the distal fragment. The axis of this fragment must be put into the right position from outside and the adjustment must not be left to the penetrating nail, as may be done safely with solid bones. With bone tumors tissue substances are frequently elected with the marrow juice which renders possible a reliable tumor diagnosis, because they are suitable for a histological examin- ation. 2£L_the_y=nai£ The pertrochanteric fracture demands a special treatment with the nail. It is a fracture which does not offer any diffi- culties with regard to reduction and fixation. Nevertheless it often causes a coxa vara position. It does, however, not snow a tendency to a delayed formation •f callus or pseudarthrosis. Moreover in nailing the special advantage of the new method i.e. that the fracture site is left untouched, is neutralized in that case. As the fracture must be exposed, occasional infect- ion of the operation wound will in most cases result in a suppur- ation of the fracture site. Nevertheless, the applying of the Y-shaped nail represents a great progress - it prevents a long confinement to bed. and facilitates the nursing. The stability of the Y-shaped nail is so great in comparison with the nail- ing of the neck of the femur that the patient may get up a corple of days after the operation* From what has said before the indication is obvious. All those patients should be nailed who are endangered by a long confinrment -to bed, and the danger of an infection must be accept- ed, The advantages and disadvantages of this method must be explained to patients not belonging to this croup, so that they may decide themselves whether the operation should be performed or not. The method of inserting the nail depends on the shape of the fracture. It will be advisable to obtain an exact knowledge of the fracture lines before the operation. It must be admitted that in many cases this may not easy. Freouently it will be possible to get the necessary clearness only by X-ray pictures made before the operation in the extension of the fracture. The fragments wedged together before are then detached, and it is easier to make out the fracture lines. The fractures of the trochanter major and its immediate neighbourhood may be suitably divided into six groups for the nailing (A. W. FISCHER and MAATZ) (ill, 89) Group I, the subtrochanteric fracture is secur'd sufficient- ly by a conical nail. The other groups should be treated with the Y-nail. The shapes of the fragments are very important for the indications and the technicue of the operation. The operation is very simple in its principal features. The marrow nail must be inserted into the femur to the supposed correct depth. The corticalis must be pierced 8 centimeters be- low the nail head end, and the transverse nail must be introduced a little into the slit of the marrow nail. The X-ray pictures to be taken then show the fracture reduction probably necessary and the correction of th« height and rotation movement still re- cuired. (ill. 90), This can be done very accura- tely, because the shift- ing of the marrow nail in an upward direction to the corticalis drill hole which must be wide- ned to the corresponding side, is executed with the chisel or rongeur. After that the marrow nail' (with the transverse nail in- serted) can be turned to the reouired angle in the marrow cavity. Further control X-ray pictures must then show that the transverse nail points to the middle of the neck of the femur in both its planes. This means that it must not be seen in the second plane, because it is covered completely by the marrow nail (ill# 91)* This technirue reouires a wide opening of the fracture site. For some patients large operation wound will cause a most injurious wound Shottk and it will, as is to be expected, be in great danger of becoming infected* Therefore it was necessary to construct a special insertion apparatus which makes it possible to insert Y-nails percutane- ously (MAATZ), it has been tested sufficiently, the nail- ing of pertrochanteric fractures with it can be strongly recom- mended. Injurious effects on the wound were more rarely observ- ed and thp wound shock is not greater than with the other methods of marrow nailing. Nevertheless, a very strict indication is to be recommended, i, e, in the beginning only patients of group II should be nailed (to whom a long confinement to bed would prov dangerous). The number of persons which can be subjected to this operation will afterwards increase nui ckl\r, The principle of the insertion apparatus s very simple. It serves merely to introduce the transverse nail percutaneous- ly into the slit of the marrow nail. It is easy to understand that the surgeon will at first feel somewhat uneasy on seeing the different drills and handles belonging to the insertion apparatus. They are, however, required and the surgeon will be fully compensated for all his trouble the moment the trans- verse nail disappears in the lateral slit in the skin, and makes its way safely through the slit in the marrow nail into the femur neck and head. ' Ne£C£int^on_of_the_inse5liipn-appfirat^3_llllJ._22l The clamp (A) must be put on the head of the marrow nail with the dovetail (B) in place. It is then fastened with the screw pin (C) which enters the hole of the nail. The nail and the screw must fit tightly. If, however, a single Y-nail is not sufficiently tight the cam (D) may be raised a little by any mechanism with a layer of metal and, if necessary, this may be removed again. Into the hole (E) a hook can be inserted and the nail extracted as reouired in each individual case. The tube (G) fixed in the tube holder (F) by a screw ser- ves as a guide for the drills and cutters. The axis of th° tube is 2,5 millimeters below the middle of the slit of the marrow nail.. This variation is allowed because • the drill or cutter touching the corticalis oblinuely is forced upwards for about 1,5 millimeter even when penetrating slowly, and a further millimeter is allowed for any further deviation (111. 93). Til A transverse nail ,touching marrow nail too far hist- ally will invariably wand er upwards i * e. into the slit, A transverse nail in- serted too far prox- imally will jam and force the marrow nail upwards, so that the previous result of the operation will be frustrated. As a further protect- ive measure the tr.:.:. r verse nail is round shaped at its upper leading edge. If, however, this nail should catch on the edge of the marrow nail its oblique part must be in» creased with a file. It is necessary that the operating surgeon should know that the axis of the tubes . is fixed in the above described way as he must ascertain r each time before inserting the nail whether the insertion apparatus is straight or not. Otherwise he may try to correct this inten- tional deviation in position. The tube (F) extends to the corticalis. Nevertheless the first hole must be made in the corticalis with a short thick cutter in order to prevent a slipping upwards. After that the rod with a cutting edge (J), which tapers towards the tip, is introduced. In order to get a better view of the control picture it is made so long that it can pene- trate the head of the femur. The final drill hole in the corticalis is made with the large cutter (K) and it must be removed again when no longer need- ed, since the cutter holder must be introduced from outside owing to lack of space. The cutter is made of such length that damage cannot be caused by the nail and the cutter coming into contact, A small bone edge is of necessity left standing for the time being at the upper edge of the bored hole, which must be penetrated by the transverse nail (ill, 94)* The transvers'e nail may be inserted easily and this' bone edge is not of any disadvantage, since the nail is pressed downward by it, i, e. in a favorable direction. The surgeon must, however, know that 1 vr: ng the insertion a certain resistance r.:ust be overcome, If it is very strong it will bo better to extract the nail again. If a faulty contact -ith the upper end of the slit in the marrow nail occurs this may be observed by a deformation of the upper edge of the transverse nail. Any,deformation must be removed accord- ingly. I U_. 1 III Cl- luM The nail or nail lor the neck of the femur (L) must be inse3~ted into the slit of the elastic fork (M) , It fits well into the tube holder (F) th>~ tube of which has been re- moved, and in this way the proper guidance of the nail is achieved. It should bo inserted with light hammer strokes. The nail must then extend out of the corticalis as far as the length of the notch in the nail holder (M)* All cutters are connected with the drill rod by an inserted piece (N; so that the connection with the shaft is possible by simply putting them on during the freouent changes. After a sufficient knowledge of the apparatus the nail- ing is easy to describe: 1, The patient is put on the table in a supine position as usual for nailing the neck of the femur. The injured leg must be left free. It must be brought to a maximal flexion, abduct- ion and inner rotation, in order to achieve thr insertion of the guide rod and the marrow nail. If the patient is pushed towards the edge of the table as far as possible the guide rod can easily be inserted percutaneously into the marrow cavity oblieuely from below , As soon as the nail has reach- ed. the marrow cavity the guide r*d can be removed, because it is not required any more, or, rather, it must be removed because the head end of the marrow nail leaves no room for it. Before the disappearance of the nail head end under the skin the clamp is put on, nail is then inserted in such a way that 2 centimeters of it stand out of the bone. It is very important that the nail should hav~ the right position from the very first. One has to go by the position of the knee-cap whrch indicates the frontal plane of the leg. The plane of the clamp must be 30° posteriorly fro the frontal plane, be- cause the neck of the femur points frontwards to the same de- gree . (111. 95) 2, The position of the patient is the same as with the femur neck nailing, 3, Final closure, 4, Fixing of the tube which indicates the direction of the incision. The tube can then be inserted, 5, The small corticalis drill hole is made with the short thick cutter (H). Push it in slowly in order to avoid too str-ong a -deviation in the upward direction, 6, Insertion of the rod with cutter tip (J), 7, X-ray pictures in two planes for the control of the location of the fracture and the position of the sound with cutting tip, (ill, 96) - 8, Make all the necessary corrections# First of all the rod with cutting tip must be removed, then the necessary correction of the S?*>***nre site must be ma.de, then the correct- ion of the height and rotation must be made accordingly# After that one can decide whether the tube must bo removed or not and a new hole in th© skin made, or, whether the elasticity of the soft parts makes th% shifting, of. it possible. If corrections are necessary the following procedure is required: ILL.'I S' Illdfc 9. A new opening is made with a gimlet and the rod with a cutter is inserted under X-ray control in two planes. 10, If the sound wire indicates that the direction of the tube leads through the middle of the neck of the femur and its head, the corticalis drill hole must be made with the large cut- ter, 11, Removal of the cutter and tube. 12, Insertion of the nail held in the nail carrier with light hammer strokes, 13, X-ray control, 14* Removal of the clamp and clesing of the wound. The short skin incisions may be closed with two stich sutures Objections raised at first against the use of the cutter without adeouate protection of the surrounding soft parts were disproved by the experiences during the operations. Kven the large cutter is ouickly covered with soft particles torn from the surrounding soft parts, so that serious wound necrosis will not occur and practice has whown that the wounds will heal most satisfactorily. If in the beginning the surgeon is not ouite certain how far the marrow nail is to be driven in and in which position the use of the sound (just as with the nailing of the nee1: of the femur) will be helpful. This must be inserted at the supposedly right place in the supposedly r ght direction. By X- ray control the necessary corrections are indicated which must be made during the later insertion of the marrow nail. The position of the clamp must be in accordance with the position of the sound , It will not be difficult to nail groups I, II and IV with the technioue described abov*, Greater difficulties are to be encountered with groups III, Va and Vb especially concerning the reduction if the axis of the neck of the femur cannot be directed towards the middle of the marrow cavity. But this fracture too cannit be nailed without reduction. Besides this is must be considered that a normal nail must be insufficient, fhese nails are short and conically shaped. They provide insufficient protection against rotation as the friction between the marrow nail and the bone is not strong enough. This m-ans that fractures of the groups III, Va and Vb are not protected against rotation for the far end of the trans- verse nail is situated either in the fracture slit, so that it 'kes not find a sufficient hold to prevent a rotation, or the second fracture line lies even farther distally; In such cases therefore marrow nails must be used which are made to measure (up ti a thickness of IS millimeters) ahd which secure sufficient friction betw/e-n the nail and the bone. It stands to reason that the insertion will be somewhat difficult due to displacement of the fracture by stroks with a hammer, especially if corrections arn to be made, and moreover, there is the difficulty of achieving better position of the nail, since the nail can be turned only after it has been almost entirely removed. Therefore, th~ greatest care must be taken in these cases that th° nail gets its right position in the distal fragment at first. At the same time the position of the knee cap must be considered. Slight rotation faults can then bo correct- ed by turning the nail with the distal fragment. Patients treated with the Y-nail may get up as scon as their general condition and the wound pain makes it possible. The fragments are well stabilized. The nail can be removed after the lapue of three to four months. For this purpose only slight skin incisions are required. The patient lies on the sound side. Then removing the nail of the neck of the femur the leg must bo rotated inwards by about 3® degrees if the operation starts at the old wound. This larg~ nail too may safely be left in the bon©, especially if the patients are elderly who do not want to undergo another operation. Patients who we re operated on three years ago have so far net suffered any injurious effect caused by this large metal- lic foreign body. 21jL_THF_SUPRAC0hDYLAR_T-FRaCTURE As a reduction and fixation of this most complicated kind of fracture is freouently not possible without an operation, and as any exposure is dang°"Ous to the knee joint and th° leg, it seemed to be advisable to utilize the advantages of marrow nail- ing with these fractures. The nail alone, however, will not suffice, since it is sure to push aside the condyles and to pene- trate into the knee joint, i.e. by itself it always will aggra- vate the situation. The shattered condyles must bo connected with each other by a transverse wire and the tip of the marrow nail must be united with this transverse nail, to prevent a slipping back of the short fragment. It would be of very great advantage*if.the transverse nail could, be inserted r.nto a hole in the nail tip, directing the drill according to tie X-ray indications. Cir- cumstances, however, make it impossible for the time being to •btain that apparatus which is indeed not difficult to construct, S« at present we must try to get along with th© following method. !♦ Insertion of a transverse wire and attaching of a Beck tension clamp. The wire should be rather near to the fracture site, in order to diminish the danger that the marrow nail may miss it* 2, Insertion of a marrow nail, which must have a wedge- shaped notch at its tip, is done in the usual way. Its exact length should be ascertained with special care* 3. In general the reduction of fracture is possible only by using strong forces. The estension wire with clamp which lies in the lower fragment, and a traction strap dn the fragment serve as handles. The required amount of traction, must be carefully judged, for putting t nail upon the transverse wire can be furnished only by a block-and.-tackle, The operation is rendered more difficult by the fact that one cannot obtain an unobstructed view (the tension clamp ob- structs the view). Nevertheless it is worth while trying it. During the first six to seven it will be advisable accord* to the position of the fracture and th° formation of callus - to apply a slight additional extension to the leg (Tuberositas tibiae or Calcaneus), for it must be borne in mind that a com- pound fracture is concerned, and that the muscular forces act- ing upon the fragments (Muscles of the leg and the thagh) are very great. The leg must be put on a Braun splint. After the removal of the nail and the transverse wire a plaster cast must be applied to prevent new displacement of the fragments. In the beginning special attention must bo paid to the in- sertion site of the transverse nail. As the final position of the stretched wire is different from the one during the insert- ion, corresponding relieving incisions are recommended, in order to prevent the wire from cutting through t skin and perhaps causing an infection. £2a_ATYPICAL_NAILINGS Atypical nailing may be necessary if the broken bone shows pathological curvatures or if a neighbouring joint shows a stiffs: ing or a contraction, There are many possibilities, some of which may be mentioned here: 1. Transverse fracture of the leg immediately below the knee-joint which is stiffened at 135°. The straight nail was inserted into the tibia from the middle of the femur passing through the knee-joint, A correction of thr* curvature of the axis to 160° was achieved. The patient, a woman, was able to walk three weeks later without any inconvenience (ill. 97) 2. A sixteen-year old patient with a stiff strongly re- curved knee-joint was subjected to osteotomy. It was performed immediately the knee-joint and stabilized in the thigh bone with a marrow nail 60 centimeters long, inserted percutaneous- iy at the trochanter, and extending to the middle of the tibia . The patient was able to get up with a straight leg a fortnight later , 3. A subtro- chanteric fracture combined with a stiff hip-joint can be nail- ed in the usual way. As regards the position of the leg it must be taken into consideratior that in this case the distal fragment must correspond to the po- sition of the proximal fragment, i.e. the leg must be extended in t’r corresponding positio n As regards the after- treatment it must be taken into considerates. that the forces which have the tendency to displace the fragments are unusually strong, since the hip-joint cannot giv~ way, Thei fore a confinement to bed for at least 4-5 weeks will be necessar During the first few weeks a comfortable po- sit i in of the patient must be achieved and renewed traumas must be prevented, since the leg is a large lever which has the power to bend the nail or even cause a piercing of the trochanter by the nail, 4* A pseudarthrosis of the arm above the elbow combined with a stiffened elbow joint can be treated with a straight thigh nail (here 10 millimeter), as it may pass through the joint. In this way the disadvantage of the long arm of the lever is neutral- ized which renders healing so difficult in such cases, 5, Oblique fracture of the leg at the border between the middle and the lower third in the case of a very nar- row cavity (diameter 6 millimeters). As the turn spread nail is insufficient and as a nail with inclined plane (diameter 6 millime- ters) may break easily, a conical nail can be insert- ed from the inner aspect of the bone. Thus the nail has a diameter of 9 millimeters at the fracture site (111. 98). 6, Another poss- ibility to apply the mar- row nail is to be found in the transplantation of bon- es. In the case shown here the proximal end of the ulna was resected because of a relapse of osteitis fibfosa. Two months later the pro- jecting end of the radius Illustr. 97 Iu ’■vc.s divided and the middle piece having become detached in this way was transplanted upon the end of the ulna. The bones were stabilized with marrow nails. Three weeks tho radius osteotomy wound was well he led and the transplanted part was well covered with callus. Nothing can be said yet about the further development. 7, An artificial ankylosis was first tried three times with a long nail. One of the patients, a tabetic, though sum- moned several times did not show up for the after-treatment. Owing to the peculiarity of the disease hi s statement that he had no trouble at all is not to be regarded as positive proof of a bony healing. second patient, a woman died one year the operation, before the af t-■ r-tr had started, v/hilr the third patient did not show any bony healing. The nail which is 60 centimeters long and must be inserted p rcutaneously at the trochanter has indeed a great stability, which however,is probably not sufficient to neutralize all forces tending to cause a lateral displacement at the bone ends, since the places at the bones of the leg and the thigh which hold the nail firmly are 25 cen- timeters apart. Along that stretch the sta- bilization nf the extremity depends entirely on the stability of the nail. Every movement in the nail and every bending motion of the nail must result in a bending movement at the bone ends, in most cases it must even lead to a lateral displacement as each tipping over would cause a, lengthening of the extremity. All the muscles, and the weight of the body, however, constantly counteract this lengthening so that there is no other possibility than f«r the fragments t» evade this by a movement resulting in a lateral dis- placement, These movements will be the stronger the more distant the points a and b are from each other (111, 99). Therefore the proper thing is to make a V-shaped resection. The bone ends will then be firm- ly pressed together just as with an oblique fract- ure, and any movements leading to a lateral displacement will be im- possible, (ill. 106) 8, If a bone has a second fracture and in the case of a refracture we have to deal with a special kind of nailing operation, In that case it must be borne in mind that after the removal of the nail, near the aid fracture site the «arr#w cavity has become rather narrow bv the formation of narrow callus. Unfortunately we have to choose a thinner nail than in the first nailing. Otherwise the nail would get stuck at the old fracture site , It will be more advisable to fit the nail tip with a few jagged teeth so that it may create for itself a new bod in the callus. IilM 9. In a single case, at the request of a patient, we prophylactically stabil- ized a thigh bone with the marrow nail in the presence of a subtrochanteric metastas- is from a breast carcenoma (ill. 101). Three months later the metastasis could not be observed any more, so that we could re- move the nail again. 10, The subtrochanteric osteotomy is performed most successful v/ith a strong- ly curved thigh nail. By the elastic transformation in its longitudinal profile the nail is fixed firmly, so that an additional plaster cast is not reruired. ILl. ! o 4 23* BIBLIOGRAPHY ASMUSSEN, Ergebn. d. Marknagelung a. d. Kieler Univ. K^inik, Inaug. Diss. Kiel 1942 BAUER, K.H,, Marknagelung oder Drahtextension? Zbl. Chir. 1943$ 254 BERNHARD, Med. Gos. Giessen 1942 BOEHLER, Vorschlag zur Marknagelung nach KUENTSCHER bei frischen Oberschenkelschussbruechen. Chirurg 1943, H. 1. - Apparate fuer die Marknagelung naeh KUENTSCHER. Chirurg 1943y H. 3. BRUECKE, Ein unangenehmer Zv/ischenfall bei der Marknagelung d. Oberschenkels. Zbl. Chir, 1943; 337 CELLARIUS, die Marknagelung von Pseudarthrosen. Zbl. Chir, 1943; 745 DENGLER, Siegfried, Kritische Betrachtungen zur operativen Be- handlung von Gelenkfrakturen Zbl, Chir. 1943; 121 EHALT, Erfahrungen mit der Marknagelung nach KUENTSCHER. Zbl, Chir. 1941; 1378 - Bisherige Ergebnisse unse"er operativen Knochenbruchbehand- lung mit besonderer Beruecksichtigung des KUENTSCHER-Nagcl: Arch, orthop. u. Unfallchir. 42. Bd., 3. H. 1943; 377 FISCHER, A.W., Chir. Kongress Berlin 1940 und Dresden 1943 - u. M a a t z, Unsere Frfahrungen mit der Marknagelung nach KUENTSCHER. A-f.ch. Klin. Chir. 203, Bd, 4. H. ; 531 (1942.) - u, R e i c h, wie steht os um die Frage der Osteomyelitis bei der Marknagelung komplizierter Frakturen? Zbl, Chir. 1943;29> GEHL, W,, Die Marknagelung nach KUENTSCHER, dargestellt an cinem ausgewaehlten Fall, Mtsschft. Unfallhoilk. 49. Jahrg., H. 12 (1942) GFRHARDT, F., Frfahrungen mit der Markbolzung Zbl. Chir. 1942; 185’ GRIESSMANN u.REICH, Vergleichende Untersuchungen uebei? den Ab- lauf der Xnochenbruchheilung bei der Marknage- lung und bei den mit Gipsverband behandelten Frakturo.n, Arch. f. klin. Chir. 205. Bd,, 3. H. 1944, S. 455-474. GUETTNER, Med. Ges. Wien 2. VII, 42 HAASE, Die KUENTSCHER Nagelung bei Spontanfraktur durch Hyper- nephrommetastase. Zbl. Chir. 1943; 1266 HAEBLFR, Die Hstabile 0steosynthese" der Knochenbrueche und ihre wirtschaftliche Bedeutung. Zbl. Chir. 1943; 374 - Die Marknagelung der herarmbrueche vom proximalen 'nde. Zbl. Chir. 1943; 734 Die Indikation zur stabilen Osteosynth^se, Monatsschrift Unfallheilk, u. Begutachtung 50, Jahr