AN AUTOMATICALLY CONTROLLED, INEXPENSIVE PUMP-OXYGENATOR Adrian Kantrowitz, M.D., Stuart Reiner, B.E.E., M.E., and Donald Abelson, M.D., Brooklyn, N.Y. Sxe the early reports of the successful development of extracorporeal cir- culation and open heart surgery,® 14 there has been widespread interest in refinement of techniques. Our group has directed its energies recently to the development of an inexpensive, self-controlled pump-oxygenator. During the 10 years that we have been interested in experimental extra- corporeal circulation, we have studied various types of pumps’? as well as bubble oxygenators, screen- and disc-type oxygenators. At present, we prefer a rotating dise oxygenator which we have modified to suit our particular needs. The apparatus consists of a rotating dise oxygenator originally suggested by Bjork’ and more recently, described by Kay and Cross” * #2! and others,?° combined with a roller-type pump. Venous blood is brought from the superior and inferior venae cavae to the oxygenator by gravity. Since a venous pump is not used, a significant source of trauma to the blood is eliminated. The pool of blood in the oxygenator serves as a venous reservoir so that the latter component is also eliminated. The discs are driven directly by a small frac- tional horse power motor at 120 revolutions per minute. Oxygenated blood leaves at the opposite end of the chamber and is carried by large-bore Tygon tubing to a roller pump. The pump then returns the blood to the arterial side through a filter. Excess blood in the opened, bypassed heart is collected by a suction tip attached to a second roller pump which delivers it to a com- bined defoaming chamber and bubble trap reservoir. From here, the blood is returned to the venous end of the oxygenator. It has been shown by McCaughan and his co-workers!® and others that most of the trauma to the blood removed from the opened heart results from excessive mixing of air with the blood in the suction pump. To minimize this effect, a control button has been placed on the suction handle which is pressed only when suction is desired. Thus the pump is turned on and off at the table, and the influx of air is minimized. From the Departments of Surgery, Maimonides Hospital of Brooklyn, and the State Uni- versity of New York-College of Medicine at Brooklyn. Supported in part by a grant (H-3023) from the National Heart Institute of the National Institutes of Health. Read at the Thirty-ninth Annual Meeting of The American Association for Thoracic Surgery at Los Angeles, Calif., April 21-23, 1959. 586 Vol. 38, No. 5 AUTOMATICALLY CONTROLLED PUMP-OXYGENATOR 587 November, 1959 The parts of the system which are in contact with blood consist of Tygon tubing, rubber, stainless steel, and silicone-coated Pyrex glass.